紧急医用混合器呼吸机Mslvm12

| 物理规格 | |
| 显示 | 10.4“液晶显示器 |
| 模式: | 气动和电控系统 |
| 工作模式: | 关闭;半封闭;半开放 |
| 手推车: | 配有4个防静电橡胶脚轮; 其中两个是可锁定的制动和容易机动性与脚踏制动设置 |
| 技术规格: | |
| 气体要求: | 医用氧气和一氧化二氮,压力范围为280〜600kpa |
| 安全阀门 | <12.5kPa |
| 呼吸频率 | 1bpm〜99bpm |
| 气体浓度输出 | 21%〜100%可调 |
| 最大分钟通风 | ≥18L/ min,允许偏差:±20% |
| 呼吸模式 | a / c,sippv,ippv,simv,psimv,cpap,pcv,peep,sign,spont,manual |
| 呼吸机合规性 | :≤30Ml/ kPa |
| 窥视: | 0〜1.0kpa |
| Ptr: | ,-1.0〜1.0kPa |
| IMV频率: | 1〜12次/ min,允许偏差:±15%。 |
| I / E比率: | 4:1~1:4 |
| 潮汐体积范围 | 0〜2000ml分辨率:10ml |
| 吸气平台: | 0~1s |
| 最大安全压力 | ≤6.0kPa |
| 叹: | 每80次控制呼吸一次深呼吸, 吸气时间为设定点的1.5倍,允许误差为-10%或15%。 |
| 压力极限范围: | 1.0〜6.0kPa |
| 气道压力报警: 听觉和视觉和 与黄色和红色颜色表示 | 上限报警:0.9 kPa〜5.4 kPa |
| 低:0.5kPa | |
| 通风量报警: | ≥18l / min |
| 电源报警 | 交流/直流电源发生故障后立即发出报警 报警时间:保持> 120s |
| 气道压力持续高于15 hPa±1 hPa持续15 s±1 s,然后机器将发出声音报警,压力将显示为红色,持续高 压力红色警报字样显示在麻醉呼吸器的屏幕上。 | |
| 运行条件 | |
| 环境温度: | 10〜40℃ |
| 相对湿度: | 不高于80% |
| 气压: | 860 hPa〜1060 hPa |
| 电力需求: | 100-120 Vac,50 / 60Hz; |
| 空气源要求: | 医用氧气和笑气,额定压力为0.3〜0.5MPa。 |
| 注意:用于麻醉机的交流电源必须良好接地。 | |
| 注意:使用的麻醉机必须配备符合ISO 9918:1993的二氧化碳监测仪, 符合ISO 7767:1997的氧气监测器和符合51.101.4.2的呼气气体体积监测器 医疗电气设备第二部分:麻醉系统的安全和基本性能的特殊要求。 | |
| 存储 | |
| 环境温度: | -15oC〜+ 50oC |
| 相对湿度: | 不高于95% |
| 气压: | 86 kPa〜106 kPa。 |
| 应储存在无腐蚀性气体,通风良好的房间内 | |
| 包 | |
| 包装盒 | 符合GB / T 15464的要求 |
| 在包装盒和产品之间, 软材料,其具有适当的厚度以防止运输期间的松动和相互摩擦 | |
| 防潮和防雨保护,以保护产品免受自然损害。 | |
| 安全报警 | |
| 气源压力 | 立即报警; “高电平报警”指示灯闪烁,发出声音报警。 |
| 气道压力报警 | 气道压力超过15hpa。最大延迟不超过17s。 “高级别报警”指示灯闪烁,发出声音报警。 |
| 潮气量报警 | 低于50ml,一次呼吸循环后报警将提高。 |
| 通风量报警 | 下限:0〜12L / Min;上升:18L / Min |
| 电池电压报警 | 它在电池电压低于10.5V时发出警告。 |
| 开机自检 | 工作前自检,确保工作正常。 |
| Physical specifications | |
| Display | 10.4'' LCD Display |
| Mode: | pneumatically powered and electrically controlled system |
| Working Mode: | Closed; Semi-Closed; Semi-Open |
| Trolley: | Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions |
| Technical Specifications: | |
| Gas requirement: | Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa |
| safety valve | ﹤12.5 kPa |
| Respiratory Rate | 1bpm~99 bpm |
| gas concentration output | 21% ~ 100% adjustable |
| Maximum Minute ventilation | ≥18L / min, permissible deviation: ±20 % |
| Breathing Mode | A/C, SIPPV,IPPV,SIMV,PSIMV,CPAP,PCV,PEEP,SIGN,SPONT, MANUAL |
| Ventilator compliance | : ≤30 Ml/kPa |
| PEEP: | 0~1.0KPa |
| Ptr: | ,-1.0~1.0 kPa |
| IMV frequency: | 1 ~ 12 times/min, permissible deviation: ±15%. |
| I/E Ratio: | 4:1~1:4 |
| Tidal Volume range | 0~2000ml resolution ratio: 10ml |
| Inspiratory Plateau: | 0~1s |
| Maximum Safety Pressure | ≤ 6.0kPa |
| Sigh: | one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point and the permissible error is -10% or 15%. |
| Pressure limit range: | 1.0~ 6.0kPa |
| Airway pressure alarm: Audible and visual and with yellow and red color indicating | upper alarm: 0.9 kPa ~ 5.4 kPa |
| lower: 0.5kPa | |
| Ventilation Volume alarm: | ≥ 18L/ MIN |
| Power Supply Alarm | Ac/dc power supply are after failing to send out alarm immediately Alarm Time: keep ﹥120s |
| The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator. | |
| Operating conditions | |
| Ambient temperature: | 10 ~ 40oC |
| Relative humidity: | no higher than 80 % |
| Atmospheric pressure: | 860 hPa ~ 1060 hPa |
| Power requirement: | 100-120 Vac, 50/60 Hz; |
| Air source requirement: | Medical oxygen and laughing gas with a rated pressure ranging from 0.3 to 0.5MPa. |
| Attention: the AC power supply used for the anesthesia machine must be well grounded. | |
| Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System. | |
| Storage | |
| Ambient temperature: | -15oC ~ +50oC |
| Relative humidity: | no higher than 95 % |
| Atmospheric pressure: | 86 kPa ~ 106 kPa. |
| It should be stored in a room without corrosive gas and well ventilated | |
| Package | |
| packaging box | comply with the requirement of GB/T 15464 |
| Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation | |
| Moisture protection and rain protection to ensure that product is protected from natural damage. | |
| Safty & Alarm | |
| Gas source pressure | Immediate alarm; 'high level alarm' indicator flickers raising an audible alarm. |
| Airway Pressure Alarm | The airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The 'high level alarm' indicator flickers raising an audible alarm. |
| Tidal Volume Alarm | Lower than 50ml, The alarm will be raised after one breathing cycle. |
| Ventilation Volume Alarm | Lower: 0~12L/Min; Up: 18L/ Min |
| Battery Voltage Alarm | It alrms when the battery voltage is lower than 10.5V. |
| Power On Self Test | Self test before working, assure it's working normally. |