Changsheng CS9912BX inverter programmable pressure tester high pressure source analyzer
CS9912BX, CS9913AX, CS9913BX is the industry's first precision variable frequency withstand voltage tester, the output AC voltage frequency range of 40.0Hz ~ 400.0Hz, resolution of 0.1Hz.The output voltage using hardware feedback, high-speed MCU control technology, Output voltage regulation rate of less than 1%, so that it can not only be used as a pressure tester, and can be used as a variable frequency high voltage source. Normal output mode and the gradient output mode so that it can be used as production line test with pressure tester, But can also be used as an analyzer.
Performance characteristics:
model | CS9912BX | CS9913AX | CS9913BX | ||||||||
ACW | Output Voltage | range | 0.050 kV to 5.000 kV | ||||||||
Precision | ± (2% of reading + 5V) | ||||||||||
Resolution | 1V | ||||||||||
Maximum output power | 100VA (5.000 kV / 20 mA) | 250VA (5.000kV / 50mA) | |||||||||
Maximum rated current | 20 mA | 50mA | |||||||||
Lower limit current range | 0 to 20mA, 0 = No judgment lower limit | 0 to 50mA, 0 = No judgment lower limit | |||||||||
Current range | 200 uA, 2 mA, 20 mA | 200 mA, 2 mA, 20 mA, 50 mA | |||||||||
Output waveform | Sine wave | ||||||||||
Output waveform distortion | ≤2% (no load or pure resistive load) | ||||||||||
Crest factor | 1.3~ 1.5 | ||||||||||
Output signal type | DDS + power amplifier | ||||||||||
Voltage rise time | 0.3s ~ 999.9s 0 = voltage rise time off | ||||||||||
testing time | 0.3s ~ 999.9s 0 = continuous testing | ||||||||||
Voltage drop time | 0.3s ~ 999.9s 0 = voltage drop time off | ||||||||||
Intervals | 0.0s to 999.9s 0 = Interval time off | ||||||||||
Output voltage mode | N mode and G mode | ||||||||||
DCW | Output Voltage | range | 0.05kV ~ 6.000kV | ||||||||
Precision | ± (2% of reading + 5V) | ||||||||||
Resolution | 1V | ||||||||||
Maximum output power | 60W (6.000 kV / 10 mA) | 300W (6.000 kV / 20 mA) | |||||||||
Maximum rated current | 10mA | 20 mA | |||||||||
Current range | 2uA, 20uA, 200uA, 2mA, 10mA | 2uA, 20uA, 200uA, 2mA, 20mA | |||||||||
Ripple factor | ≤5% | ||||||||||
Discharge time | ≤ 200ms | ||||||||||
Maximum charge current | 20 mA | 50mA | |||||||||
Voltage rise time | 0.3 to 999.9s 0 = Voltage rise time off (except CS9913AX) | ||||||||||
testing time | 0.3 to 999.9s 0 = Continuous test (except CS9913AX) | ||||||||||
Voltage drop time | 0.3 to 999.9s 0 = Voltage fall time off (except CS9913AX) | ||||||||||
Intervals | 0.0 to 999.9s 0 = Interval time off (except CS9913AX) | ||||||||||
Delayed alarm time | 0.3 to 999.9s 0 = Delayed alarm time off (except CS9913AX) | ||||||||||
Output voltage mode | N mode, G mode (except CS9913AX) | ||||||||||
Electricity Pressure table | range | 0.050 kV - 5.000 kV (AC) 0.050 kV - 6.000 kV (DC) | |||||||||
Precision | ± (2% of reading + 5V) | ||||||||||
Resolution | 1V | ||||||||||
Displays the value | Root mean square value | ||||||||||
Electricity flow table | measuring range | AC | 0 to 20 mA | 0 to 50 mA | |||||||
DC | 0 to 10 mA | 0 to 20 mA | |||||||||
Resolution | AC | 200uA Stall: 0.1uA, 2mA Stall: 1uA, 20mA Stall: 10uA, 50mA: 10uA | |||||||||
DC | 2uA file: 0.001uA, 20uA file: 0.01uA, 200uA file: 0.1uA, 2mA file: 1uA, 20mA: 10uA | 2uA file: 0.001uA, 20uA file: 0.01uA, 200uA file: 0.1uA, 2mA file: 1uA, 20mA: 10uA 50mA stalls: 10uA | |||||||||
measurement accuracy | ≥ 2mA for ± (2% + 5 words), <2mA for ± (3% + 5 words) | ||||||||||
Offset function | The current in the test leads and accessories can be subtracted. | ||||||||||
Test mode | GND mode: RETURN Termination housing FLOAT mode: RETURN not connected to the chassis | ||||||||||
meter Time Device | range | 0~ 999.9s | |||||||||
Resolution | 0.1s | ||||||||||