Product parameters:

Model: Color-Eye 7000A
Brand: GretagMacbeth
Origin: United States
Advantage: double the speed of light, D/8 ° integrating sphere design; Can measure the reflection/transmission rate; Meet the highest color test requirements.
Order number: 20319
Features
-GretagMacbeth Color-Eye 7000A added automation features to improve operational efficiency. 7000A 70,001 and optical structure, so that both instruments measured the color data are in good agreement.
Programmed regulator-7000A has two customers. Specular reflection of a motor-driven plate can easily detect specular or diffuse color data changes. In addition, it can automatically adjust UV content and nature of D65 light source.
-Unique touch-screen displays the status of instruments and instrument settings you can easily. And to use any software that works with 7000A immediately. If you are using gelingda maikebeisi color or color formula software quality management software, you can choose to use devices or software to change the status of the instrument.
-Sample holding device can be moved accurately, including thin film measurements of diffusion or direct transmittance the sample readings. It is very important for multiple applications for opaque samples. 7000A used with gelingda maikebeisi ROPALETTE color matching software, will be able to get all the transparency color of the precise formula.
-7000A has two analyzers, single shine or flashing several times to measure samples and internal parameters as needed. This advanced design ensures long time stability and faster reading, flow.
-High concentrations of XE can accurately read the dark color or high concentration readings. In addition, more signal to noise ratio it eliminate all ambient light effects. D65 light source of the short amount of time can extend equipment life, and does not generate heat, without preheating. 7000A color with the addition of pulse Xenon light source. According to a preset time spacing, auto shine to balance two spectrometers. This continued regular adjustments can compensate for temperature variations, current, preventing electronic drift.
Product specifications
-R (white): 0.01RMS e CIE Lab
-Consistency between the instruments (LAV): 0.08 average CIE Lab
-Light source: pulsed xenon arc
-Number of flashes/measurements (normal use): 1
-Measurement time: ﹤ 1 seconds
-Spectral range: 360nm to 700nm
-Wavelength accuracy: 0.1nm (400-700nm)
-Wavelength accuracy: 0.05nm (400-700nm)
-Wavelength interval: 10nm
-Reflectance range: 0% to 200%
-Resolution: 0.01%
-Measuring aperture
Large measuring aperture (LAV) 1' (2.54cm)
Measurement aperture (MAV) 0.6' (1.5cm)
Measurement of small aperture (SAV) 0.3'x0.4' (0.75cmx1cm)
Measurement of ultra small aperture (VSAV), 0.12'x0.31' (0.3cmx0.8cm)
-Optical geometry: diffuse/8 ° (lighting/measure)
-6' (15.2cm) integration integrating sphere
-Two optical spectrum analyzer
-Transmission: direct and scattering
-Overall dimensions: length 28' width 11' height 15.25'
□71.1cm×27.9cm×38.7cm
-Operating temperature: 15 c ~32 c
-Power consumption: 36 w when no operation, continuous use 100 Watts
-Power requirements: 117~230 VAC/50~60 kHz