Mattel MT280 multi-mode thickness gage | Metal thickness gauge | Ultrasonic Thickness gauge

Mattel MT280 multi-mode thickness gage | Metal thickness gauge | Ultrasonic Thickness gauge

5800 CNY (844.5USD)

add to cart

Product parameters:

  • Customized processing: no
  • Type: ultrasonic thickness gauge
  • Brand: MiTeCH/Mattel
  • Model: MT280
  • Measuring range: 0.65~ 600


Mattel MT280 high precision ultrasonic thickness gauges

Measuring range: Launch-echo mode (0.65~ 600) mm (steel, probe subparagraph applicable); echo-echo mode (3~ 100) mm (steel, P5EE probe)

Indicating accuracy: 0.05mm (10MM); (± 0.5%H+0.01) mm (﹥ 10MM); H is the measured thickness

Communications stores: Can store 20 Group (per group of up to 100 measurement values) thickness measurement data communication interface USB1.1

Standby time: More than 50 hours (when the backlight turns off)

GPS equipment weight: 345g

Total weight: 1.5KG

Overall dimensions: 132mm×76mm×32 mm

MT280 ultrasonic thickness gauge based on ultrasonic measuring principle, may be of metal, plastic, ceramic, glass and many other good conductors of ultrasonic thickness measurement of materials, also can realize the counter measure of material sound velocity. Sealed metal housing design, defensible in harsh environment pollution, dust; Compared with the traditional methods of measurement, ultrasonic thickness gauge's advantages is that it just contacts the workpiece side to complete the measurement, its unique performance for non-destructive testing of closed pipes, containers such as thickness detection limits of working conditions provided the perfect solution. Both penetrating coating mode and normal mode MT280 dual function, measured before having to dispose of surface coating, can be measured directly. Are widely used in petroleum, chemical, metallurgy, shipbuilding, aviation, aerospace and other manufacturing equipment in the field of piping and pressure vessel corrosion thinning levels of monitoring, also can be used for all kinds of sheet metal and machined parts to make precise measurements. Is to improve the production efficiency and yield, cost savings necessary professional precision instrument.

Functional use

Suitable for measuring metals (such as steel, cast iron, aluminum, copper), plastics, ceramics, glass, fiber glass and conductor of any other ultrasonic thickness
Has two modes of thickness: launch-echo and echo-echo mode, can penetrate the coating thickness measurement and are not included in the coating thickness
Probe calibration of the zero point, two-point calibration, the system automatically correct errors
Thickness can be against measuring the sound velocity is known to improve measurement accuracy
Can be pre-set thickness value, minimum, out of range alarm
With coupling function
EL backlight display to facilitate use in dimly lit environments
Remaining battery indicators, real-time display battery remaining capacity
Has auto sleep, energy-saving features such as automatic shutdown
With a USB interface, you can easily and quickly Exchange data with the PC machine
Can be equipped with a variety of different frequency and double crystal probes for different wafer sizes
Optional equipped with computer software, transmission measurement, measurement of storage management, statistical analysis of measured values, rich features such as printing the measured value reports
Sealed aluminum housing, small, portable, reliable, suitable for harsh operating environments, anti-vibration, shock and electromagnetic interference

Working principle

Ultrasonic Thickness gauges for thickness measurement, is generated by the probes ultrasonic pulse through the coupling agent to be measured, part of the bottom surface of the ultrasonic signal is an object reflects, probes from the echo reflected from the bottom surface to be measured, accurately calculate the round trip time of ultrasonic, and according to the following formula to calculate the thickness value and the calculated results are displayed.
H=v×t/2
In the formula: h-measuring thickness
V-material velocity
T-in ultrasonic testing of round trip travel time
To ensure that the probes work, use a coupling agent to contact the sensor surface and the surface to be measured between the air and the fluid coupling used between the probe and the workpieces, known as coupling agent

Working conditions

Ambient temperature: operation temperature:-20 ℃ ~ + 50 ℃

Storage temperature:-30 ℃ ~ + 70 c

Relative humidity 90%
Environment no strong vibration, does not have a strong magnetic field, no corrosion and serious dust

Characteristics of the instrument

Suitable for measurement of metals (such as steel, cast iron, aluminium, copper, etc), plastic, ceramics, glass, fiber glass and any other ultrasonic thickness of the conductors sealed metal housing design, small size, high reliability, suitable for harsh operating environments, anti-vibration, shock and electromagnetic interference
128x64 Chinese dot-matrix LCD display, with EL backlight display to facilitate use in dimly lit environments
Both thickness mode: launch-echo and echo-echo mode, can penetrate the coating thickness measurement and are not included in the coating thickness
Measuring range than similar products, normal mode up to 0.65-600mm, penetrating coating mode 3-100mm
Mass storage, low power consumption design, can standby for several months or more of known thickness can be against measuring the sound velocity, meet a wide range of material thickness inspection needs
Come with USB data processing software can be connected to computer for data analysis, storage, and printing
Can be compatible with a variety of different frequency, different chip sizes probe
Probe calibration of the zero point, two-point calibration, the system automatically correct errors
Composite chips equipped with narrow pulse high-precision probe, blind spot is small, measuring precision
Has auto sleep, energy-saving features such as automatic shutdown, the remaining battery indicator function

Operation methods and points for attention

Operation method

The probe plug into the sensor socket; Press the switch key, along with the power on beep, instrument and the thickness of the screen displays the current mode: 'P-E' indicates that a device is working to launch-echo mode; 'E-E' indicates that a device is working in ECHO-ECHO mode. Pause automatically after entering the measuring interfaces, and displays the current set of sound values in the instrument, the parameters of the instrument parameters used for the last time before shutdown; To shut down, simply press the switch button. When shut down, thickness gauge to it both in the use of memory all setting parameters. The instrument also has an auto power off feature to conserve battery usage. If the beta-thickness gauge 5 minutes without any operation, it will automatically shut down.
Set the thickness mode: the instrument with multi-mode thickness function: launch-echo mode (P-E) and echo-Echo (E-E) mode. ECHO-ECHO mode, the instrument displays the measured value for automatic removal of the thickness of the coating layer. Inspectors often detection with coating or coating materials, such as pipes and containers. Before the test, the coating generally requires clear for thickness, otherwise it can lead to large errors. When using this instrument, if you select echo mode and with a narrow pulse-echo thickness probe (model: P5EE), can be made directly with each other without clear coating thickness. Press two thickness mode switching can be achieved. On the instrument display, when the ' + ' icon appears, Instrumentation in ECHO-ECHO mode; ' + ' Icon is not displayed, Instrumentation in launch-echo mode. 1), press the OK button in the instrument's built-in switch between sound values; Or press the up arrow and down arrow keys to change the speed of sound; 2) by pressing the CAL key instrument parameters to adjust the end state.

Measurement considerations

When measuring only a couple icons and stability, is a good measure;
If the surface to be measured when there is a lot of coupling agent, when the probe when they leave the surface to be measured, measuring error caused by coupling agent. At the end of a measurement, should quickly remove probe body surface being measured.
Acrylic resin surface probes, redrawing is very sensitive to rough surfaces, so should tap in use; When measuring the rough surface, minimizing probe paddling on the work surface.
Measurements at room temperature, the surface to be measured shall not exceed 60 c, otherwise the probe cannot be used again.

Instrument care and maintenance

Instruments should be avoided and the probe has shaken the
Avoid equipment placed too wet environments
When you plug the probe should hold activities coat along the axis, and not rotate the probe to avoid damaging the probe cable wire
Oil, dust attached to the probe line is aging, break, cable should be cleared after use of dirt