Product parameters:
- Processing customization: No
- Type: high pressure safety tester
- Brand: Changsheng
- Model: CS9915AX
- Measuring range: 0.500kV ~ 9.999kV
- Input voltage: 220
- Output voltage: 0.500kV ~ 10.00kV
- Ambient temperature: See data
- Weight: See details
- Dimensions: See details
- Voltage measurement error: ± (2% of reading +5 V)
Changsheng genuine CS9915AX program-controlled ultra-high pressure safety tester
CS9915AX series is designed for high-voltage devices designed ultra-high voltage tester, is the first domestic output voltage up to 10kV program-controlled safety tester. The output voltage using hardware feedback, high-speed MCU control technology, the output voltage regulation rate range of 1 %, To avoid instability due to input power supply voltage and load fluctuations caused by the output voltage instability test results are not accurate.Use DDS + amplifier output sine wave voltage, output voltage waveform is pure, distortion is small, the measurement results more accurate and reliable .
Features:
◆ Security anti-power wall and safety protection system can protect the safety of the operator to ensure that the test pieces are not damaged.
◆ Generated by the DDS standard sine wave, linear power amplifier driver output, the output voltage waveform is pure, distortion is small, better than 2%.
◆ fast voltage build-up time:
(1) AC output voltage of the establishment of the time for the 126ms (2) between steps and the minimum time of 450ms.
◆ With 0.4ms fast cut off function.
◆ AC withstand voltage test current minimum resolution of 0.1uA, DC voltage test current minimum resolution of 0.001uA; is the only industry with low current resolution of the tester.
◆ Output voltage has normal mode and gradient mode; normal mode is to set the multi-step output voltage test, the previous test time after the end of the output voltage from the set value is reduced to 0 in the next output of the output voltage. Refers to the set of non-output voltage test, the previous step after the end of the test time, the output voltage does not drop to 0, but according to the next set of voltage directly to the set voltage output is the only industry with two Output voltage mode of the tester.
◆ Current measurement terminal can choose to ground mode or floating mode.Grounding mode refers to the current measurement terminal and chassis connected together, this model is mainly used in the test system.Floating mode refers to the current measurement terminal suspension, not with the machine Shell connected together, this model is applied to high precision measurement occasions.
◆ can automatically save the test results in the tester, easy to check the test results; also can be connected with the RS232 interface to test the test results into the EXCEL table for data statistics, analysis.
◆ Set the voltage rise time, test time, voltage drop time and interval time.
◆ PLC interface test signal can output 24V level, can also output contacts.
◆ With the list display function, when the test step is greater than 2, after the test is over, the tester can automatically display the results of each step of the test.
◆ Arc detection with current mode and level mode two.
model | CS9915AX | CS9916AX | CS9917AX |
ACW | Output Voltage | range | 0.500 kV to 9.999 kV |
Precision | ± (2% of reading + 5V) |
Resolution | 1V |
Maximum output power | 100VA | 200VA | 500VA |
Maximum rated current | 10mA | 20 mA | 50mA |
Lower limit current range | 0 to 10 mA | 0 to 20 mA | 0 to 50 mA |
Current range | 200 uA, 2 mA, 10 mA | 200 uA, 2 mA, 20 mA | 200 mA, 2 mA, 20 mA, 50 mA |
Output waveform | Sine wave |
Output waveform distortion | ≤2% (no load or pure resistive load) |
Crest factor | 1.3~ 1.5 |
Output signal type | DDS + power amplifier |
Voltage rise time | 0.3s ~ 999.9s 0 = voltage rise time off |
testing time | 0.3s ~ 999.9s 0 = continuous testing |
Voltage drop time | 0.3s ~ 999.9s 0 = voltage drop time off |
Intervals | 0.0s to 999.9s 0 = Interval time off |
Output voltage mode | N mode and G mode |
Electricity Pressure table | range | 0.500 kV to 10.00 kV |
Precision | ± (2% of reading + 5V) |
Resolution | 2.5V |
Displays the value | Root mean square value |
Electricity flow table | range | AC | 0 to 10 mA | 0 to 20 mA | 0 to 50 mA |
Resolution | AC | 200uA Stall: 0.1uA, 2mA Stall: 1uA, 10mA, 20mA Stall: 10uA; 50mA Stall: 10uA |
measurement accuracy | ≥ 2mA for ± (2% + 5 words), <2mA for ± (3% + 5 words) |
Offset function | The current in the test leads and accessories can be subtracted. |
Test mode | GND mode: RETURN Termination housing FLOAT mode: RETURN not connected to the chassis |
meter Time Device | range | 0~ 999.9s |
Resolution | 0.1s |
Precision | ± (0.1% + 50ms) |