Product parameters:
Chang Sheng CS9916BX ultra high pressure controlled Tester analysis of breakdown of insulating materials
CS9916BX and CS9917BX are designed for high voltage devices and ultra high pressure tester, the output voltage up to 10kV. Using full-wave rectifier output voltage, output voltage ripple coefficient; The output voltage can be changed by pressing the small, can be used as high voltage devices and breakdown analysis of high-voltage insulating materials used. Using hardware to feedback, high-speed MCU control technology, the output voltage regulator rate is within the range of 1%, the measurement result reliable, stable. Meet the IEC61730-2 of the test voltage (voltage of the system 4 times +2000V) requirements.
Performance characteristics:
1, 240*64 drawing mode LCD display, user-friendly interface design is easy to operate.
2, which is in the English interface, you can meet the different needs of different users.
3, constant voltage output: output voltage regulation in the range ± 1%, avoid the input supply voltage is not stable and load changes in the output voltage changes, measurement results are not accurate.
4, with rapid discharge function, measuring instrument in DC after testing within 0.2s to be tested and the loop ended to ensure operator safety.
5, with DC test open circuit detection can prevent DC test because the test off miscarriages.
6, zero startup may prevent damage from being tested.
7, voltage gradient goes up, looking for analysis of breakdown point.
8, the current upper limit lower limit alarm and alarm functions can be set to prevent the test line off miscarriages.
9, and the current offset function eliminates leakage current influence on the test results of the test fixture.
10 in the testing process, you can manually change the size of the output voltage.
11, with ARC detection test.
12, have GFI protection.
13, which can be preset to save the test parameters: set 30 files, each file can be set to 99 test step.
14, keypad lock, protection test parameters are not modified by operators.
15, front panel software calibration, calibration can be carried out without opening arguments.
16, PLC interface standard, optional RS232C, RS485, USB and GPIB interface.
Model | CS9916BX | CS9917BX | |||
DCW | Output voltage | Range | 0.500kV~ 9.999kV | ||
Precision | ± (2% read value +5V) | ||||
Resolution | 1V | ||||
Maximum output power | 100VA | 200VA | |||
Maximum rated current | 10mA | 20mA | |||
Minimum current range | 0~ 10mA | 0~ 20mA | |||
Current gear | 2uA, 20uA200uA, 2mA, 10mA | 2uA, 20uA200uA, 2mA, 20mA | |||
Ripple | Less than 5% | ||||
Maximum charge current | 10mA | 20mA | |||
Voltage rise time | 0.3s~ 999.9s 0 = voltage rise time | ||||
Test time | 0.3s~ 999.9s 0 = continuous testing | ||||
Voltage drops | 0.3s~ 999.9s 0 = voltage drop time | ||||
Interval time | 0.0s~ 999.9s 0 = interval time | ||||
Delay alarm time | 0.3s~ alarm 999.9s 0 = delay time | ||||
Output voltage mode | N model, g-mode | ||||
Voltage meter | Range | 0.500kV~ 10.00kV | |||
Precision | ± (2% read value +5V) | ||||
Resolution | 2.5V | ||||
Show values | Root mean square value | ||||
Ammeter | Range | DC | 0 ~ 10mA | 0 ~ 20mA | |
Resolution | DC | 2uA files: 0.001uA, 20uA file: 0.01uA,200uA files: 0.1uA, 2mA files: 1uA,10mA, 20mA: 10uA | |||
Measurement accuracy | 2mA to ± (2%+5 characters), ﹤ 2mA ± (3%+5 characters) | ||||
Offset function | The current testing lines and accessories can be subtracted. | ||||
Test mode | GND models: RETURN Terminal chassis FLOAT mode: RETURN-end case | ||||
Timer | Range | 0~ 999.9s | |||
Resolution | 0.1s | ||||
Precision | ± (0.1%+50ms) | ||||