Product parameters:
Chang Sheng CS9914AX programmable precision voltage Tester test Analyzer production line high voltage source
CS9914AX/9914BX is the safety industry's first precision variable frequency withstand voltage Tester, the output AC voltage frequency range 40-400Hz resolution 0.1Hz . Hardware output voltage feedback high speed MCU Control, output voltage regulator range 1%Or less, making it not only can be used as a voltage tester to use, and can be used as inverter high voltage source. Normal output mode output mode and gradient so that it can be used as a production line for inspection of high voltage testing device, and can also make it as an Analyzer. Maximum transformer capacity is 500VA,To meet the UL1004In the 23.4.2The transformer capacity requirements.
Performance characteristics:
1, 240*64 drawing mode LCD display, user-friendly interface design is easy to operate.
2, which is in the English interface, you can meet the different needs of different users.
3, constant voltage output: output voltage regulation in the range ± 1%, avoid the input supply voltage is not stable and load changes in the output voltage changes, measurement results are not accurate.
4, with rapid discharge function, measuring instrument in DC after testing within 0.2s to be tested and the loop ended to ensure operator safety.
5, with DC test open circuit detection can prevent DC test because the test off miscarriages.
6, zero startup may prevent damage from being tested.
7, voltage gradient goes up, looking for analysis of breakdown point.
8, the current upper limit lower limit alarm and alarm functions can be set to prevent the test line off miscarriages.
9, and the current offset function eliminates leakage current influence on the test results of the test fixture.
10 in the testing process, you can manually change the size of the output voltage.
11, with real current measurement function.
12, with ARC detection test.
13, have GFI protection.
14, can be pre-set save the test parameters: set 30 files, each file can be set to 99 test step.
15, keypad lock, protection test parameters are not modified by operators.
16, front panel software calibration, calibration can be carried out without opening arguments.
17, PLC interface standard, optional RS232C, RS485, USB and GPIB interface.
Model | CS9914AX | CS9914BX | |||||
ACW | Output voltage | Range | 0.050kV~ 5.000kV | ||||
Precision | ± (2% read value +5V) | ||||||
Resolution | 1V | ||||||
Maximum output power | 500VA (5.000kV/100mA) | ||||||
Maximum rated current | 100mA | ||||||
Minimum current range | 0~ 100mA,0= no lower limit judgment | ||||||
Current gear | 200uA, 2mA, 20mA, 100mA | ||||||
Output waveform | Sine wave | ||||||
Output waveform distortion | ≤ 2% (no load or pure resistance load) | ||||||
Crest factor | 1.3~ 1.5 | ||||||
Type of output signal | DDS+ amplifier | ||||||
Voltage rise time | 0.3s~ 999.9s 0 = voltage rise time | ||||||
Test time | 0.3s~ 999.9s 0 = continuous testing | ||||||
Voltage drops | 0.3s~ 999.9s 0 = voltage drop time | ||||||
Interval time | 0.0s~ 999.9s 0 = interval time | ||||||
Output voltage mode | N model, g-mode | ||||||
DCW | Output voltage | Range | - | 0.050kV~ 6.000kV | |||
Precision | - | ± (2% read value +5V) | |||||
Resolution | - | 1V | |||||
Maximum output power | - | 300W (6.000kV/50mA) | |||||
Maximum rated current | - | 50mA | |||||
Current gear | 2uA, 20uA, 200uA, 2mA, 20mA, 50mA | ||||||
Ripple | - | Less than 5% | |||||
Discharge time | - | ≤200ms | |||||
Maximum charge current | - | 50mA | |||||
Voltage rise time | 0.3s~ 999.9s 0 = voltage rise time (except CS9914AX) | ||||||
Test time | 0.3s~ 999.9s 0 = continuous testing (except CS9914AX) | ||||||
Voltage drops | 0.3s~ 999.9s time 0 = voltage drop off (except for CS9914AX) | ||||||
Interval time | 0.3s~ 999.9s 0 = interval time (except CS9914AX) | ||||||
Delay alarm time | 0.0s~ 999.9s 0 = delay alarm time off (except for CS9914AX) | ||||||
Output voltage mode | N model, g-mode | ||||||
Voltage meter | Range | AC | 0.050kV~ 5.000kV | ||||
DC |
| 0.050kV~ 6.000kV | |||||
Precision | ± (2% read value +5V) | ||||||
Resolution | 1V | ||||||
Show values | Root mean square value | ||||||
Ammeter | Measuring range | AC | 0 ~ 100mA | 0 ~ 100mA | |||
DC | --- | 0 ~ 50mA | |||||
Resolution | AC | 200uA files: 0.1uA, 2mA: 1uA, 20mA, 100mA: 10uA | |||||
DC | 2uA files: 0.001uA, 20uA file: 0.01uA,200uA files: 0.1uA, 2mA files: 1uA,20mA, 50mA: 10uA | ||||||
Measurement accuracy | 2mA to ± (2%+5 characters), ﹤ 2mA ± (3%+5 characters) | ||||||
Offset function | The current testing lines and accessories can be subtracted. | ||||||
Test mode | GND models: RETURN Terminal chassis FLOAT mode: RETURN-end case | ||||||
Timer | Range | 0~ 999.9s | |||||
Resolution | 0.1s | ||||||
Precision | ± (0.1%+50ms) | ||||||