Product parameters:
Changsheng new CS9912BS PV voltage Tester
This series program resistance pressure test instrument are used high-speed MCU and mass digital circuit design of high performance of Ann rules test instrument, its output voltage of size, output voltage of rose, declined, output voltage of frequency completely by MCU control, can real-time displayed breakdown current value and voltage value, and has soft body calibration function, equipped with PLC interface, RS232C, and can matching RS485 interface, USB device, USB Host, GPIB, Can be easily with computer or PLC integrated test system. The ability to quickly and accurately to household appliances, instruments, lighting, electric heating appliance, computer, information safety integrated measurement machine.
Performance characteristics:
Model | CS9911AS | CS9911BS | CS9912AS | CS9912BS | ||||||||
ACW | Output voltage | Range | (0.050~ 5.000)kV | |||||||||
| Precision | ± (2% read value +5V) | |||||||||||
| Resolution | 1V | |||||||||||
| Maximum output power | 75W | 100W | ||||||||||
| Maximum rated current | 15mA | 20mA | ||||||||||
Current gear | 2mA, 15mA | 2mA, 20mA | ||||||||||
| Current limit | (0~ 15.00) lower limit mA,0= judgement | (0~ 20.00) lower limit mA,0= judgement | ||||||||||
| Output waveform | Sine wave | |||||||||||
| Output waveform distortion | ≤ 2% (no load or pure resistance load) | |||||||||||
| Crest factor | 1.3~ 1.5 | |||||||||||
| Output waveform | DDS+ linear power amplifier | |||||||||||
ACW | Voltage rise time | 0.3s~ 999.9s 0 = voltage rise time | ||||||||||
| Test time | 0.3s~ 999.9s 0 = continuous testing | |||||||||||
| Voltage drops | 0.3s~ 999.9s 0 = voltage drop time | |||||||||||
| Interval time | 0.3s~ 999.9s 0 = interval time | |||||||||||
DCW | Output voltage | Range | (0.050~ 6.000)kV | (0.050~ 6.000)kV | ||||||||
| Precision | ± (2% read value +5V) | ± (2% read value +5V) | ||||||||||
| Resolution | 1V | 1V | ||||||||||
| Maximum output power | 30W | 60W | ||||||||||
| Maximum rated current | 5.000mA | 9.999mA | ||||||||||
| Current gear | 2mA, 5mA | 2mA, 10mA | ||||||||||
| Current limit | (0~ 5.000) lower limit mA,0= judgement | (0~ 9.999) lower limit mA,0= judgement | ||||||||||
| Ripple | ≤5% (6kV/5mA) | ≤5% (6kV/10mA) | ||||||||||
| Discharge time | ≤200ms | ≤200ms | ||||||||||
| Maximum charge current | 5mA | 9.999mA | ||||||||||
| Voltage rise time | 0.3~ 999.9s 0 = voltage rise time | 0.3~ 999.9s 0 = voltage rise time | ||||||||||
| Test time | 0.3~ 999.9s 0 = continuous testing | 0.3~ 999.9s 0 = continuous testing | ||||||||||
| Voltage drops | 0.3~ 999.9s 0 = voltage drop time | 0.3~ 999.9s 0 = voltage drop time | ||||||||||
| Interval time | 0.3~ 999.9s 0 = interval time | 0.3~ 999.9s 0 = interval time | ||||||||||
Electricity Pressure Table | Range | 0.050kV~ 6.000kV | ||||||||||
| Precision | ± (2% read value +5V) | |||||||||||
| Resolution | 1V | |||||||||||
| Show values | Root mean square value | |||||||||||
Electricity Flow Table | Measuring range | AC | 0 ~ 20mA | |||||||||
DC | 0 ~ 10mA | |||||||||||
Resolution | AC | 2mA files: 1uA,20 (15mA) mA, 10uA | ||||||||||
DC | 2mA files 1uA,10 (5mA) mA, 10uA | |||||||||||
| Precision | ± (2%+5 characters) | |||||||||||
Offset current | 0 ~ 2mA | |||||||||||
Timer | Range | 0.1s~ 999.9s | ||||||||||
Accuracy/resolution | ±1%/0.1s | |||||||||||