Product parameters:
Nanjing Changsheng CS9913AS AC and DC withstanding voltage tester
model | CS9913AS | CS9913BS | CS9914AS | CS9914BS | ||
ACW | Output Voltage | range | 0.050 kV to 5.000 kV | |||
Precision | ± (1% of reading + 5V) Voltage ≥1.000kV ± (2% of reading + 5V) Voltage <1.000kV | |||||
Resolution | 1V volts / Step | |||||
Maximum output power | 250VA (5.000kV / 50mA) | 500VA (5.000kV / 100mA) | ||||
Maximum rated current | 50mA | 100 mA | ||||
Lower limit current range | 0 to 50mA, 0 = No judgment lower limit | 0 to 100mA, 0 = No judgment lower limit | ||||
Current range | 200 mA, 2 mA, 20 mA, 50 mA | 200 mA, 2 mA, 20 mA, 50 mA, 100 mA | ||||
Output waveform | Sine wave | |||||
Output waveform distortion | ≤2% (no load or pure resistive load) | |||||
Voltage rise time | 0.5s to 999.9s 0 = voltage rise time off | |||||
testing time | 0.1s ~ 999.9s 0 = continuous testing | |||||
Voltage drop time | 0.5s to 999.9s 0 = voltage fall time off | |||||
Intervals | 0.1s ~ 999.9s 0 = When the test time is reached, the tester will stop automatically | |||||
Arc detection | (0.00-20.00) mAac / (0-9) 0 = off | |||||
Arc detection time | (0.0-999.9) seconds 0 = Disable | |||||
Output voltage mode | N mode and G mode | |||||
model | CS9913BS | CS9914BS | ||||
DCW | Output Voltage | range | 0.05kV ~ 6.000kV | |||
Precision | ± (2% of reading + 5V) | |||||
Resolution | 1V | |||||
Maximum output power | 120W (6.000 kV / 20 mA) | 300W (6.000 kV / 50 mA) | ||||
Maximum rated current | 20 mA | 50mA | ||||
Current range | 2uA, 20uA, 200uA, 2mA, 20mA | 2uA, 20uA, 200uA, 2mA, 50mA | ||||
Ripple factor | ≤5% | |||||
Discharge time | ≤ 200ms | |||||
Maximum charge current | 20 mA | 50mA | ||||
Voltage rise time | 0.3 to 999.9s 0 = voltage rise time off | |||||
testing time | 0.3 ~ 999.9s 0 = continuous testing | |||||
Voltage drop time | 0.3 ~ 999.9s 0 = voltage drop time off | |||||
Intervals | 0.0 to 999.9s 0 = Interval time off | |||||
Delayed alarm time | 0.3 ~ 999.9s 0 = Delay alarm time off | |||||
Output voltage mode | N mode and G mode | |||||
Arc detection | (0.00-10.00) mAac / (0-9) 0 = off | |||||
Arc detection time | (0.0-999.9) seconds 0 = Disable | |||||
model | CS9913AS | CS9913BS | CS9914AS | CS9914BS | ||
Electricity Pressure table | range | 0.050 kV - 5.000 kV (AC) 0.050 kV - 6.000 kV (DC) | ||||
Precision | ± (1% of reading + 5V) Voltage ≥1.000Kv ± (2% of reading + 5V) Voltage <1.000kV | |||||
Resolution | 1V | |||||
Displays the value | Root mean square value | |||||
Electricity flow table | measuring range | AC | 0 to 50 mA | 0 to 100 mA | ||
DC | 0 to 20 mA | 0 to 50 mA | ||||
Resolution | AC | Lt; 2 mA is 0.001 mA and & gt; 2 mA is 0.01 mA | ||||
DC | Lt; 2 mA is 0.001 mA and & gt; 2 mA is 0.01 mA | |||||
measurement accuracy | ± (2% of reading + 2 words) | |||||
Offset function | The current in the test leads and accessories can be subtracted. | |||||
Test mode | GND mode: RETURN Termination housing FLOAT mode: RETURN not connected to the chassis | |||||
meter Time Device | range | 0~ 999.9s | ||||
Resolution | 0.1s | |||||
Precision | ± (0.1% + 50ms) | |||||
Memory group / test step | 50 groups / 40 steps | |||||