Product parameters:
- Customized processing: no
- Brand: meet macros
- Model: YT-GM
Product introduction:
GM gloss measuring instrument is widely used in papermaking, printing, paints, coatings and other industries, used to determine the surface gloss. The geometry of the instrument, spectral and photometric characteristics in line with the relevant standards.
75 ° gloss measurement method is mainly used in coating paper and paperboard, can also be used for uncoated paper and paperboard, or a low printing gloss paper and paperboard printing;
20 ° gloss measurement method is mainly used in cast coated paper, wax paper, polishing paper, high gloss paper and Board, also apply to higher printing gloss paper and paperboard printing and high
Gloss ink painting;
45 ° gloss measurement methods applicable to the determination of aluminum foil paper, metalised paper and metal composite paper and paperboard;
60 ° gloss measurement law mainly applicable to all kinds of paper, coating, paint and so on.
Second, GM gloss meter standard:
GB/T 8941: paper and Board-determination of specular gloss (20 ° 45 ° 75 °);
GB/T 9754: paints and varnishes paints not containing metallic pigments film of 20 °, 60 ° and 85 ° determination of specular gloss;
ISO 2813: paints and varnishes-20 °, 60 ° and 85 ° non-metals determination of specular gloss of paint film;
Third, GM gloss meter specification and main technical parameters:
1, and CIE D65 lighting Visual function, a light linear: 0.3%;
2 measurement range: 75 ° angle gloss: 1~ 378 (gloss units);
60 degree gloss: 1~ 999.5 (gloss units);
45 ° angle gloss: 1~ 509.4 (gloss units);
20 degree gloss: ~ 2037 (gloss units);
3, repeatability: 0.5 gloss units;
4, accuracy: 1 gloss unit;
5, the sample size (length x width); mm: 50×50;
6 power: 220V ± 10%,50Hz;
7, the working environment: temperature 0~ 35 ℃, relative humidity does not exceed 85;
8, overall dimensions (length x width x height); mm: 290×350×160;
9, quality: 7kg.
Four GM gloss meter product performance characteristics:
Instruments using optical, mechanical, electrical integration of microcomputer measurement and control technology, with testing data statistic processing function to print output, measuring surface gloss
Degrees.
1. the equipment has a variety of incident angle (75 °, 20 °, 60 °/45 °);
2. with a gloss of polished black glass as standard. Refractive index of 1.567 and has ideally smooth surface of black glass, gloss set at 100 (gloss units)
3. range extension technology, the instrument can measure any surface (lowest zero gloss to the ideal mirror) finish.
4 the instrument with power off memory function, and can automatically print a test report. Computer memory, zero adjustment, sample number, statistical computing, digital display and print reports log
Value, average value and statistical value results.