JFL-BZ20°60°85° Smart gloss meter user manual
A reference standard and scope
JFL-BZ20°60°85• Intelligent gloss meter is in compliance with ISO-2813Standards of high-precision three-point integrated Intelligent
Gloss meter, suitable for coating , Paint , Ink, Plastic , Stone , Floor , Wood furniture Surface specular gloss units
(Gs ) Measurements.
Second, the instrument specifications
Instruments and indicators to meet the national metrological JJG 696First degree gloss meter standard.
1Display range: 0.0-300.0(Gloss units) (G s)
2Indication error: ﹤ 1.2Gloss units (g s)
3Stability: 0.4Gloss units (Gs)/30Here minutes
4Measure the window size: 16×48 (mm2)
5Instrument size: 138×36×82 (mm3)
Third, the instrument use:
Press the switch, the instrument in measurement mode, instrument placed on the sample, display values for the samples under the angle of light
Gloss value (g s) . When you press the 'Angle /Shift ' Key ,Switch between different angle measurement ,Display lights corresponding angle. No matter that
Corner
Measurement, when the measured value is greater than 300.0Gloss units, display 'O v-e-r '.
When the low battery display 'bAtt ', At this point in time with the special charger charging charging time is four hours left
Right.-charging standard '??'
Calibration:
When displayed on the instrument in the standard numerical angle when the difference between the nominal value corresponding to the Standards Board should be calibrated as follows:
Instruments to measure the window in Black wool Board ,Press 'Reset number /Zero ' Key ,Displays a value of0.0.
Standard plate black velvet with black glass panel, measuring window, Position the equipment Center ' triangle mark groove edge Center
Press the ' calibration /Test ' button, go to settings, display '000.0' And the first '0' Flashing, press the ' view /Move
' Key flashes right, press ' reset number /Zero ' key ' Flash ' Value from the 0-9-0Changed by pressing the ' view /Shift '
With the ' reset number /Zero ' display value corresponding to the black glass standard nominal values are the same, then press ' calibration /Testing ' to finish
Set. And instruments in the white ceramic standard plate, Position the equipment Center ' triangle mark groove edge Center , Show values
The corresponding nominal value with the white ceramic standard plate error shall not exceed ±1.2Gloss units, you should take into account standards Board dirty q
Problem. which angle error exceeds in whichever mode should be calibrated in accordance with the above procedure. You can then sample Jin
Line measurement.
IV Note:
This set of four rechargeable batteries in the instrument,
Package includes:
Host a single, standard set of (black and white plates with a zero Board (standard box on black velvet)) ,Wiping paper a copy of instructions
A copy of the book, a copy of the certificate, charger, case is one.