Product parameters:
Product description:
This product is designed and manufactured in accordance with the basic principles of emission spectrum analysis equipment. Widely used in agricultural fertilizers, soil analysis, analysis of cements, silicate industry analysis and
Measurement and other aspects of the clinical laboratory and pathology research and detection of psychiatric patients taking lithium salts.
Product features:
● K, Na Can also test also tests for lithium.
● LCD Display, direct digital readout.
● With a primary function and switches off automatically to protect the flame size device.
● Simple, fast, and less of the sample.
● Using liquefied petroleum gas as fuel Gas.
Technical parameters:
Model | FP640 |
Consistency: | Standard solution with continuous sample introduction, 15s Relative maximum value of equipment in change ≤ 3% Per minute test 1 Times to determine 6 Instrument indication of relative maximum variation ≤15% |
Repeatability: | Repeat on the same standard solution 7 Successive independent measurement ≤ 3% |
Linearity error: | K: ≤0.005mmol/L (0.0100~ 0.0800) mmol/L Na : ≤0.03mmol/L(0.0500~ 0.400)mmol/L |
Limit of detection: | K: Detection limit ≤0.004mmol/L ; Na : Limit of detection ≤0.008mmol/L |
Response time: | Response time< 8S |
Sample injection volume: | Sample injection volume < 6mL/min |
Complete set: | Host one; Compressed air compressor units; Attachments material |
Normal working conditions: | Ambient temperature: 10℃~ 35℃ |