Product parameters:
Main features:
Dot matrix (with blue backlight) LCD display, full Chinese operation interface
Device, conductivity, TDS, salinity, and temperature measurement
Automatic temperature compensation, automatic calibration, automatic range switching power off protection function
The test results can be stored, delete, read, and print up to store 200 measuring data and provides instant printing, storage
Print two copies of the print mode for user to select
With calibration function, you can set the electrode constant or TDS conversion factor
RS-232 interface can be connected with the TP-16 printer, mine-use REX DC1.0 magnetic data acquisition software to communicate with your computer
Use titanium alloy electrode conductivity cell constant 0.01cm-1 and seal measuring Groove, suitable for high water measurements
Equipped with the factory-DJS-1C conductivity (PT black), T-818-B-6 type electrode
Equipment levels: level 1
1, measuring range:
Electrical conductivity: (0~ 1.999x105) Μ S/cm divided into six-speed range automatically switches
( 0~ 1.999 ) μS/cm; ( 2.00~ 19.99 ) μS/cm;
( 20.0~ 199.9 ) μS/cm ( 200~ 1999 ) μS /cm
( 2.00~ 19.99 )mS/cm ( 20.0~ 199.9 )mS/cm;
TDS: (0~ 19900) mg/L is divided into five-speed range, you can automatically convert
( 0~ 10.00 ) mg/L: (10.0~ 100.0 ) mg/L
( 100~ 1000 ) mg/L ( 1.0~ 10.0 ) g/L
( 10.0~ 19.99 ) g/L;
Salinity: salt mass percentage (0.00~ 8.00)
Temperature: ( -5.0~ 105.0) ℃
2, basic error:
Electrical conductivity: ± 0.5% (FS) ± 1 Word
T d s: ± 0.5% (FS) ± 1 Word
Salinity: 0.1
Temperature: ± 0.3 ° c ± 1 Word
3, stability: ± 0.3% (FS) ± 1 Word/3h
4, temperature compensation range: (0~ 50) ¡æ
5, power: Xinying YX-965K universal power supply (9V DC, 300mA is negative)
6, dimensions (mm): 290x200x70
7, weight: about 1 kg
8, constants and the corresponding electrode conductivity measuring range:
Electrode constant cm-1 | 0.01 | 0.1 | 1.0 | 10.0 |
Measuring range | 0.000~ 19.99μS/cm | 2.00~ 199.9μS/cm | 20.0μS/cm~ 19.9ms/cm | 200μS/cm~ 199.9ms/cm |