Product parameters:
LX-A Shaw Hardness tester Device determination in supporting the production of the same types of load racks, rubber in the laboratory, plastic hardness standard test piece to determine the standards. Portable hardness meter head, rubber mounted in the device (plastic) objects to determine the surface hardness.
Note: the hardness of plastic film and plastic hardness testing. Plastic sample and adjust in accordance with GB/T2918-1998 of the plastic sample conditioning and testing related to the standard regulations. Product execution country standard GB/T531-1999 the Pocket hardness test method for indentation hardness of rubber, this standard is equivalent to international standards of the Rubber-Determination of indentation hardness by means of taking into account the JJG304-2003 of the pocdet Har-dness meter of the type a durometer calibration procedures in the relevant provisions.
Features:
Pointer indicates, easy reading, accuracy, precision
Portable design, easy to carry, hand measurements can also be mounted on a matching test, test
Parameters:
Dial: 0 ~ 100HA
Dropper on the trip: 2.5 mm
Recommended measuring range: 10 ~ 90HA
Needle size: ф 0.79 mm
Overall dimensions: 115*60*25mm
Weight: ~ 0.5 kg








Method of use: | |
Put sample placed in substantial of plane Shang, hold hardness meter, pressure foot in the hole of pressure needle distance try block edge at least 12mm, smooth to put pressure foot pressure in sample Shang, cannot has any vibration, and keep pressure foot parallel Yu sample surface, to makes pressure needle vertical to pressure into sample, by imposed of force to just enough to makes pressure foot and sample completely contact, except another has provides, must in pressure foot and sample completely by touch Hou 1 seconds within readings, if is other interval time readings is must description, In samples from different locations at least 6mm measured hardness value 5 times and take the average. | |
Terms of use: | |
⑴, determination of hardness should be checked before the pointer should point to zero-in a free State. Pointer when the deviation from zero, you can loosen clamping screw in the upper-right, turn, zero. And hardness on the glass plate, the pressure pin pressure foot surface in close contact at the time of the glass, the pointer should point to 100+/-0.5HA, such as when no points to 100+/-0.5HA, slightly press the pressure pin several times, still not saying when 100+/-0.5HA, this hardness cannot be used. When used in shore hardness test rack, pull the handle, raise the table to load weight lifting, which is pin closely contact with the foot flat on the glass desk, the pointer should point to 100+/-0.5HA. If not 100
| |