AT573 DC bias current source | Test level 1 v | Precision 0.05%

AT573 DC bias current source | Test level 1 v | Precision 0.05%

1+
5680 CNY (827.03USD)
3+
5600 CNY (815.38USD)
5+
5500 CNY (800.82USD)

add to cart

Product parameters:

  • Customized processing: no
  • Type: digital current measurement instruments
  • Brand: meet macros
  • Model: AT573
  • Measuring range: -10A-10A


AT573 DC bias current source

AT573 DC bias current source (DC magnetization power bias current source) is used for inductors and transformer windings inductance measurement applying DC bias current, this device apart from the need to achieve the required current, also small impact on the measurement of inductance, wide frequency response (superimposed DC current can be used when high-frequency LCR Meter measurement) and other requirements. AT573 DC bias current source with color vacuum VFD Display, controlled by microprocessor, working parameters automatically memory. Automatic current balanced principles to achieve the stability of AC-DC superposition of current, constant current output with forward or reverse 0-10A range. Additional errors the product, frequency response to be increased substantially, greatly expanded a magnetic core inductor applications in bias test range. Directly controlled by Amber AT2818,AT2816A,AT2816B,AT2817A,AT2817,AT810A digital bridges and more current scan output functions, provide a good solution for the analysis of magnetic materials.

Technical specifications
Current range -10A-10A (200mA/2A/10A
Current reading accuracy ± 1% ± 3 words
Current resolution 200mA ranges: 0.1 mA,
2 a ranges: 0.1 mA,
10A-range: 10 mA
Current output type single point output, more output shared output
Frequency response 50 Hz-200kHz
L ﹤ 2 k ω resistance range ω and l ﹤ 8/I (h), output current I
Output voltage 0-8 v
Basic impedance accuracy ≤ 1%
Scanning mode manually, automatic (Flyback)
Points 2-15 points
Set delay time 0-99 seconds
Warm-up time 20 minutes
Performance characteristics
Four color vacuum fluorescent display VFD
Direct current is controlled by the AT2818,AT2816A,AT2816B,AT2817A,AT2817,AT810A automatic or manual scanning output
Long time high current output
User data automatically saved
Frequency response range: 50Hz-20KHZ low additional error, wide frequency response range
Single point, multi-point scanning, equal parts scanned three current output
0-99s delay time setting
RS232 interface
Shock protection, prevent damage to LCR tables
Single output current reverse switch, adjustable current
Desirable/shared scans, with Flyback function, manual/auto scan
Working parameters automatically memory
Operating 0~40℃, ﹥ internal overheating temperature 70 ℃, humidity 20~80%RH