Product parameters:


Function | 234 wire grounding resistance measurement of soil resistivity; Ground voltage, AC voltage measurement |
Power supply | DC 9V (R14S 1.5V 6 zinc manganese dry battery section, continuous standby 300 hours) |
Range | Ground resistance: 0.00 ~ 30.00k ω ω |
Soil resistivity: 0.00 m~ ω 9000k ω m | |
Measurement mode | Precise 4-wire, 3-wire measurement, simple 2-wire measurement of grounding resistance |
Methods of measurement | Ground resistance: rated current pole method, measuring current: 20mA Max Soil resistivity: quadrupole law (Wenner method) Voltage: average value rectifier (P(S)-ES Interface) |
Testing frequency | 128Hz/111Hz/105Hz/94Hz(AFC) |
Short-circuit test current | AC 20mA max |
Open the test voltage | AC 40V max |
Test voltage waveform | Sine wave |
Range of electrode spacing | Set 1M~ 100m |
Shift | Ground resistance: 0.00 ~ 30.00k fully automatic shifting |
Soil resistivity: 0.00 ω m~ 9000k m fully automatic shifting | |
Display mode | 4King size LCD display, control white screen backlight |
Measuring instructions | Measurement of LED flashing, LCD count down, the progress bar indicates |
LCD size | 128mm×75mm; LCD display field: 124mmx67mm |
Meter size | High thickness: 215mmx190mmx95mm |
Standard test line | Section 4: Red 20m, black 20m, yellow 10m, green 10m 1 |
Simple testing line | Article 2: the Red 1.6m, black 1.6m 1 |
Auxiliary grounding rods | 4: φ 10mmx150mm |
Measuring time | Voltage to ground: about 3 times per second |
Ground resistance, soil resistivity: about 5 seconds/second | |
Measurement frequency | More than 5,000 times (short circuit test, measuring 1, stop for 30 seconds then again) |
Line voltage | Measurement of AC 600V |
RS232 interface | With RS232 interface, software control, stores can upload data into the computer, save the print |
Communication line | RS232 communication line 1, 1.5M |
Data storage | 300 groups 'MEM ' Storage instructions, showFULL ' Symbol indicates that storage is full |
Data retention | Keep data 'HOLD ' Symbol indicates |
Data access | When you read data 'READ ' Symbol indicates |
Overflow display | Overrange overflow 'OL ' Symbol indicates |
Interference test | Automatic identification of interfering signals, interference voltage higher than 5V 'NOISE ' Symbol indicates |
Auxiliary ground test | With auxiliary grounding resistance test function, 0.00K Ω~ 30kΩ(100R+rC ﹤50k Ω, 100R+rP ﹤50k Ω) |
Alarm function | When the measured value exceeds the alarm set point, the ' beep – beep – beep--' alarm |
Battery voltage | When the battery voltage drops to about 7.5VWhen the battery low symbol display, reminder to replace the battery |
Power consumption | Standby: 20mA (backlight off) |
Power the backlight turns off: 45mA (backlight off 25mA) | |
Measurements: approximately 100mA (backlight off) | |
Quality | 4.5kg (including packaging) |
Working temperature and humidity | -10 ℃ ~ 40 ℃; 80%RH the following |
Storage temperature and humidity | -20 ℃ ~ 60 ℃; 70%RH the following |
Overload protection | Grounding resistance measurement: C(H)-E , P(S)-ES Each port AC 280V/3 seconds |
Insulation resistance | 20MΩOr more (500V between circuits and shell) |
Pressure resistance | AC 3700V/rms (between circuits and housing) |
Electromagnetic properties | IEC61326(EMC) |
For safety | IEC61010-1 (CAT III 300V, CAT IV 150V, pollution degree 2); IEC61010-031; IEC61557-1 (resistance); IEC61557-5 (dist); JJG 366-2004. |
Supplied accessories | Instrumentation: 1; Grounding pin: 4; Test line: 4; Simple test line: 2; RS232 communication cable: 1; Discs: 1; Enclosures: 1 PC |