Product parameters:

  • Brand: MiTeCH/Mattel
  • Model: MH320
  • Product type: Leeb hardness tester
  • Measuring range: (170-960), HLD, (17-68.5), HRC, (19-651), HB,
  • Test force: 10kg
  • Overall dimensions: 212mmx80mmx32mm
  • Customized processing: no


First, an overview:

Mattel MH320Portable Leeb hardness tester based on the hardness measurement, can be measured easily and quickly on a variety of materials, hardness measurement values are displayed at the same time, can be prepared in a variety of hardness (Brinell HB , Rockwell HRC , Vickers HV , Shaw HS ) Free conversion allows for preset tolerance limit, out of range alarm. Host the product design and integrated high speed thermal printer, imported high-speed core support on-site instant printing capabilities of the printer, imported long-lasting paper data storage for a long time. Low-power design concept based on stable IC ICS, standard high capacity, long life, high speed rechargeable Ni-MH battery pack. Chinese display, menu-driven operation is simple and convenient, seven special impact device without calibration, automatic identification, USB Interface transfer, mass storage, and PC Personalized data processing software.

Second, the technical parameters:

  • Measuring range: (170-960) HLD,(17-68.5)HRC,(19-651)HB,
  • Measuring direction: supports vertically downward, diagonal downward, horizontal, diagonal up, straight up
  • Material: steel and cast steel, alloy tool steel, stainless steel, grey cast iron, nodular cast iron, cast aluminum alloy, copper-zinc alloys (brass), copper-Tin alloys (bronzes), pure copper, forged steel
  • Hardness standard: Richter (HL)Blagojevich (HB), Rockwell B/C(HRB/C), Rockwell A(HRA), Vickers (HV), Shaw (HS)
  • Display: dot matrix LCD , 128×64Graphic dot matrix LCD
  • Data storage: Max 500Groups (number of shocks32~ 1)
  • Thermal printer, working quietly, you can print any number of copies of test results, meet the needs of field
  • Print width: (57.5±0.5) mm ; Printer paper roll diameter: 30mm
  • Working voltage: 6VNi-MH battery pack ; Power supply: 9V/500mA ; Charging time 2.5~ 3.5Hour
  • Working time: about 150Hours (does not print, when the backlight turns off) ;Interface standard: USB1.1

Third, the applicable materials:

Steel and cast steel, alloy tool steel, stainless steel, grey cast iron, nodular cast iron, cast aluminum alloy, copper-zinc alloys (brass), copper-Tin alloys (bronzes), pure copper, forged steel

Four, working conditions:

Ambient temperature: operation temperature:- 10~ + 50; Storage temperature: -30℃~ + 60; Relative humidity90% ;

Environment no strong vibration, does not have a strong magnetic field, no corrosion and serious dust.

Five features:

  • According to Leeb hardness measurement principle can be carried out on a variety of materials testing ;Software calibration.
  • A host can be equipped with 7Different impact devices, automatically identifies impact devices types ,Replacement without recalibration.
  • Support ' forged steel (Steel ) ' Material, when used D/DC Impact device test ' forged ' sample can be directly read HB Without manual look-up table.
  • On the hardness value can be pre-set limit, out of range alarm, convenient batch testing needs.
  • Use of large-screen 128×64Graphic dot matrix liquid crystal display, information-rich and intuitive.
  • Chinese display, menu-driven operation, the operation is simple and convenient. Have EL A backlit display, convenient for use in dimly lit environments.
  • Maximum storage 500Groups (number of shocks 32~ 1) Hardness measurement of data, each set of data including single measurement values, average, measurement date, impact direction, frequency, material, hardness and other information.
  • Thermal printer and instrument integration as one, quiet, fast print speeds, you can print test reports.
  • Built-in rechargeable Ni-MH rechargeable batteries and control circuits; Continuous work is not less than 150Hour; Has auto sleep, energy-saving features such as automatic shutdown.
  • The remaining battery indicator icon on the LCD, real-time display battery life; Indicates the charging process, the operator can keep informed of doc.
  • Have USB Interface, you can easily, quickly and PC For data exchange. Can be equipped with the software, transferring measurement results, measured value storage management, statistical analysis of measured values, print a measurement report, rich features such as bulk set the instruments parameters to meet the higher demands on quality assurance and management.
  • Instrument small and portable integrated thermal printer can print, overall dimensions: 212mm×80mm×32mm

Six instrument configuration:

Serial number

The name

The number of

Notes

Standard configuration

1

Instruments hosting

1Taiwan

2

DImpact device

1Nondescript animals

3

Standard hardness blocks

1Piece

4

Nylon brush

1Nondescript animals

5

Small support rings

1Nondescript animals

6

Power adapter (charger)

1Nondescript animals

7

Printer paper

1Volume

8

Random data

1Copy

9

ABS Instrument case

1Nondescript animals

Select configuration

10

Nylon brush b

GImpact device use

11

Special impact device and supporting ring

See Schedule 3 and Schedule 4

12

Data transfer and analysis software CD

1Set

Application on your computer

13

Communication cable

1Long things

Special impact device

Seven, power on display and a brief description:

Hardness standard

Average

The impact direction

Battery

The number of shocks

Measuring value

Materials

Battery: displays the remaining capacity.

• Impact: the impact direction.

• Average value: after the set number of shocks, displays the average appeared.

• Standard hardness: hardness of the current measured value.

• Measurements: current single measurement values (average tip-free), the current average (mean).

• Display: indicates more than conversion or measurement range, lower conversion or measurement range.

• Materials: materials for the current setting.

• Number of shocks: impact number displayed when the measurement has been completed, with the shortcut key display settings when setting the impact and number of number of number of shocks, when browsing a single measurement values display the corresponding number of single measurement.

Eight USB data processing software:

• Deposit: the measured data is stored

• Out: data from the host to bring up

• Print: print the data output, the A4 interface

• Overload: overload displayed when the set tolerance limits

• Clear: clear data

• Connections: setting and host connections

• Disconnect: disconnect set from the host

• Download: download the data to a local, realization of mass storage

• Settings: parameters setting

• Help: instant help feature

Seven, the MH310 MH320 differences are as follows:

1, On the show: the MH310 is a segment LCD, economy function is relatively simple. MH320 menu-driven operation, powerful, easy to use 2, storage: MH310 can store up to 100 Group (impact and number of 32~ 1), MH320 storage 500 group.

3, On the standard: MH310 data processing software for optional configuration, MH320 standard standard data processing software.

Eight notes:

  • This can only be provided by the use of the company's special battery and power adapter (charger), as it may cause equipment damage, battery leakage, fire or explosion.
  • When printing or when printing is completed please do not open the paper cover, contact printer core by hand or any part of the body, so as to avoid excessive temperatures result in Burns.
  • Do not put batteries into fire, do not short circuit the battery, broken up or heat, as it may lead to battery leakage, fire or explosion.
  • If stored for a long time the instrument (not used), please make sure the battery switch is in the 'OFF' position, as it may lead to battery failure or leakage and damage to the printer.

IX, warranty description:

Under normal operating conditions of the user in accordance with the instructions, product quality problem in two years, the manufacturer warranty. Wearing, except consumables.

Ten tables measuring range:

Materials

Hardness

Impact device

D/DC

D+15

C

G

E

DL

Steel and cast steel

Steel and cast steel

HRC

17.9~ 68.5

19.3~ 67.9

20.0~ 69.5

22.4~ 70.7

20.6~ 68.2

HRB

59.6~ 99.6

47.7~ 99.9

37.0~ 99.9

HRA

59.1~ 85.8

61.7~ 88.0

HB

127~ 651

80~ 638

80~ 683

90~ 646

83~ 663

81~ 646

HV

83~ 976

80~ 937

80~ 996

84~ 1042

80~ 950

HS

32.2~ 99.5

33.3~ 99.3

31.8~ 102.1

35.8~ 102.6

30.6~ 96.8

Steel Forged steel

HB

143~ 650

CWT , ST

Alloy tool steel

HRC

20.4~ 67.1

19.8~ 68.2

20.7~ 68.2

22.6~ 70.2

HV

80~ 898

80~ 935

100~ 941

82~ 1009

Stainless steel

Stainless steel

HRB

46.5~ 101.7

HB

85~ 655

HV

85~ 802

GC. IRON

Gray cast iron

HRC

HB

93~ 334

92~ 326

HV

NC , IRON

Nodular cast iron

HRC

HB

131~ 387

127~ 364

HV

C. ALUM

Casting of aluminum alloy

HB

19~ 164

23~ 210

32~ 168

HRB

23.8~ 84.6

22.7~ 85.0

23.8~ 85.5

BRASS

Copper-zinc alloy

(Brass)

HB

40~ 173

HRB

13.5~ 95.3

BRONZE Copper-tin alloy

(Bronze)

HB

60~ 290

COPPER Pure copper

HB

45~ 315

11, special impact device technical data sheets and apply:

Special impact device

DC(D)/DL

D+15

C

G

E

Impact energy

Impact body weight

11mJ

5.5g/7.2g

11mJ

7.8g

2.7mJ

3.0g

90mJ

20.0g

11mJ

5.5g

Ball hardness:

Ball diameter:

Ball head material:

1600HV

3mm

Tungsten carbide

1600HV

3mm

Tungsten carbide

1600HV

3mm

Tungsten carbide

1600HV

5mm

Tungsten carbide

5000HV

3mm

Diamond

Impact device diameter:

Impact device length:

Impact device weight:

20mm

86(147)/ 75mm

50g

20mm

162mm

80g

20mm

141mm

75g

30mm

254mm

250g

20mm

155mm

80g

Specimens maximum hardness

940HV

940HV

1000HV

650HB

1200HV

Surface roughness Ra of specimen:

1.6μm

1.6μm

0.4μm

6.3μm

1.6μm

The minimum weight of the specimen:

Direct measurement

Need stable support

Need compact coupling

﹥5kg

2~ 5kg

0.05~ 2kg

﹥5kg

2~ 5kg

0.05~ 2kg

﹥1.5kg

0.5~ 1.5kg

0.02~ 0.5kg

﹥15kg

5~ 15kg

0.5~ 5kg

﹥5kg

2~ 5kg

0.05~ 2kg

Specimen minimum thickness

Compact coupling

Minimum depth of hardening layer

5mm

≥0.8mm

5mm

≥0.8mm

1mm

≥0.2mm

10mm

≥1.2mm

5mm

≥0.8mm

Ball indentation size

Hardness 300HV

Indentation diameter

Indentation depth

0.54mm

24μm

0.54mm

24μm

0.38mm

12μm

1.03mm

53μm

0.54mm

24μm

Hardness 600HV

Indentation diameter

Indentation depth

0.54mm

17μm

0.54mm

17μm

0.32mm

8μm

0.90mm

41μm

0.54mm

17μm

Hardness 800HV

Indentation diameter

Indentation depth

0.35mm

10μm

0.35mm

10μm

0.35mm

7μm

--

--

0.35mm

10μm

Scope impact devices

DC Measurement hole or cylinder barrel;DL Measurement of thin narrow slots or holes ; Type d for General measurements

D+15Small contact surface, long, suitable for measuring Groove or concave surfaces

CImpact is small, on the surface damage is very small, without destroying the hardened layer, suitable for measuring small lightweight components and surface hardening.

GMeasurement of large thick and its surface is rough castings and forgings

EMeasurement of high hardness material