Product parameters:
LD-4Low-speed centrifuges in hospital laboratories, clinics, laboratory of Biochemistry, qualitative analysis of plasma, serum urea and vaccine manufacturing.
Two Technical parameters:
1Power supply: 220V 50HZ
2Host speed: 4000r/min
3Maximum relative centrifugal force: 2800g
4Capacity: 4×100ml Or 12×10ml
5Noise: ﹤75Db(A)
6, Weight: 49㎏
Three Safety precautions:
Prohibited do not have turn under no-load operation; Running should inspect the head for any corrosion damage before; When turned to run do not touch running head; Is prohibited in turn load not balanced State operations.
Four Instructions for use:
Centrifuge the Workbench should be flat and solid, workplaces should be neat and clean, dry and well ventilated, the ambient temperature 5-32℃Is appropriate. Open the centrifuge Chamber and turned around and wipe clean. Load weighing test tube, close the centrifuge lid, speed adjustment knob, rose to the required speed. Before each stop, speed control knob on the minimum position must be. Cut off the power. Wipe the inside and the back, close the centrifuge lid.
Five Maintenance:
Do not place any objects on the centrifuge cover. Every time after use, be sure to wipe the cavity clean and back, centrifuge cover cover prevents dust, dirt. Centrifuge, if not used in a long time again before using centrifuge lid should be open for some time, drying chamber. Motor carbon brushes in use 3000Hours later, in time Replacement to avoid wear and commutator. Centrifuge with long-term use, natural wear is normal. In one year , If fails or in case of problems, the company will continue to provide quality service to assist.