Product parameters:
It can be used for the determination of dissolved oxygen in water bodies, such as tap water monitoring, aquaculture farm, environmental protection, sewage treatment plant, beverage industry and scientific research units etc. The instrument is novel in shape and easy to carry, suitable for field and field operation. Long, can be used as a routine laboratory analysis equipment.
The main technical parameters
① Measuring range
Dissolved oxygen concentration: 0.00~ 19.99 (Mg / L)
Dissolved oxygen saturation: 0.0~ 199.9 (% )
temperature: 0.0~ 40.0 (° C).
② Automatic temperature compensation range: 0.0~ 40.0 (° C)
③ Salinity Calibration Range: (0.0~ 50.0) G / L
④ Basic error of electronic unit
Dissolved oxygen concentration: ±0.10 mg / L;
Dissolved oxygen saturation: ±2.5% ;
Temperature: ±0.3℃ ±1.
⑤ Basic error of instrument
Dissolved oxygen concentration: ±0.50 mg / L ( Calibration water temperature and the measured water temperature difference should be <15Lt; 0 & gt; C )
Dissolved oxygen saturation: ±10.0% ( Calibration water temperature and the measured water temperature difference should be <15Lt; 0 & gt; C )
temperature : ±0.5℃ ±1Words
⑥ Electronic unit stability: ±0.10 mg / L / 3h
⑦ Instrument stability: ±0.20 mg / L / 1h
⑧ Instrument repeatability: 0.20 mg / L
⑨ Response time: not more than 45s ( 20℃ 90%response )
⑩ Residual current: not more than 0.10 mg / L.
Dimensions: 180Respectively70Respectively35 (Mm ), Weight: 250g