Medical Equipment Baby Warmer H-1000 Infant Incubator

Medical Equipment Baby Warmer H-1000 Infant Incubator

900-1000 USD

add to cart
Model NO. H-1000
Certification CE
Group Infant
Medical Devices Reg./Record No. Gsyjxz20123540587
Temperature Rise Time ≤30min30min0min
Temperature Uniformity ≤1°ce;1°cg;ce;1°c
Trademark Healicom
Origin China
Type Infant Incubator
LCD Display Without LCD Display
Medical Device Regulatory Type Type 2
Temperature Control Mode Air Control
Temperature Variability ±0.5°csmn;0.5°cg;csmn;0.5°c
Air Temperature 20°c~37°cc~37°cg;cc~37°c
Specification ISO
HS Code 90189090
Standard Composition:
1. Servo Control Mode: air temp;
2. Instant Parameter Displaying: set temp, air temp and heat power rate;
3. Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air temp override, fan failure, system failure; failure; failure;
4. ﹥37°C Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;unction;tion;
5. Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf;helf;helf;
6. RS-232 Output Connector (Use for network system )
Specification:
Power Model H-1000-1E 110 / 120 V ac, 50Hz / 60Hz
Model H-1000-2E 220 / 240 V ac, 50Hz / 60Hz
Power Input ≤400VA00VA00VA
Refer to IEC60601-1, this equipment belongs to ClassI, Type BF
Its operation is continuous
Alarm Power Alarm Power Failure: Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle
Fan Alarm Fan Failure: Activates if Fan motor malfunctions
Sensor Alarm Air Sensor Failure: Air Sensor malfunction, refer unit to service
Deviation Alarm High Air Temperature: Activated if displayed Air Temperature is 15°C above set pointpointpoint
Low Air Temperature: Activated if displayed Air Temperature is 25°C below set pointpointpoint
Over Temperature Alarm Over Air Temperature: Activated when displayed Air Temperature was approach 38°C for set temperature ≤37°C, or approach 40°C for set temperature ﹥37°C in Air Modegt;37°C in Air Modeegt;37°C in Air Mode
System Alarm System Failure: Activated if Temperature Control System malfunctions
Temperature Control Range and Correlative Specifications Air Temperature 20°C~37°C,for set temperature ≤37°C, 37°C~39°C, for set temperature ﹥37°Cor set temperature >37°Ceg;Cor set temperature >37°C
Temperature Rise Time ≤30min0min0min
Temperature Variability ±0.5°Cmn;0.5°Ceg;Cmn;0.5°C
Temperature Uniformity ≤1°C;1°Ceg;C;1°C
Environment Requirement Environment Temperature Operation Temperature Range +20°C~+30°C+30°Ceg;C+30°C
Store Temperature Range -40°C~+55°C+55°Ceg;C+55°C
Environment Humidity Operation Humidity Range 30%~75%
Store Humidity Range 10%~100%
Atmospheric Pressure Range 700hPa~1060hPa
Other
Specifications
Noise Level Within Hood Environment ≤55dB(Environment noise below 45dB)5dB)5dB)
Carbon Dioxide (CO2) Level within the Hood Less than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress
Air Velocity Over Mattress less than 0.3m/s
Infant Bed Max Load 10kg
Transfusion Shelf Max Load 2kg
Cabinet Max Load 6kg
Function of 'SILENCE' KeyE" Keyt; Key Make the alarm to be silence for 10 minutes by press this key Deviation Alarm
Sensor Alarm
Fan Alarm
Factory Default Settings Temperature Control Mode Air Temperature Mode
Air Set Temperature: 32°Cdeg;Cdeg;C
Temperature Units:°Cdeg;Cdeg;C

NOTE: Open Access Panel or Access Door or the use of supplies or other equipment within the Incubator, which can alter the air flow pattern, may effect temperature uniformity, temperature variability, the correlation of the Incubator temperature reading to center mattress temperature and skin temperature.