| Model NO. | H-800 |
| Certification | CE |
| Group | Infant |
| Medical Devices Reg./Record No. | Gsyjxz20123540587 |
| Temperature Rise Time | ≤30min30min0min |
| Temperature Uniformity | ≤1°ce;1°cg;ce;1°c |
| Trademark | Healicom |
| Origin | China |
| Type | Infant Incubator |
| LCD Display | Without LCD Display |
| Medical Device Regulatory Type | Type 2 |
| Temperature Control Mode | Air Control |
| Temperature Variability | ±0.5°csmn;0.5°cg;csmn;0.5°c |
| Air Temperature | 20°c~37°cc~37°cg;cc~37°c |
| Specification | ISO |
| HS Code | 90189090 |
| Power | Model H-800-1E | 110 / 120 V ac, 50Hz / 60Hz | |||
| Model H-800-2E | 220 / 240 V ac, 50Hz / 60Hz | ||||
| Power Input | ≤400VA00VA00VA | ||||
| Refer to IEC60601-1, this equipment belongs to ClassI, Type BF Its operation is continuous | |||||
| Alarm | Power Alarm | Power Failure: Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle | |||
| Fan Alarm | Fan Failure: Activates if Fan motor malfunctions | ||||
| Sensor Alarm | Air Sensor Failure: Air Sensor malfunction, refer unit to service | ||||
| Deviation Alarm | High Air Temperature: Activated if displayed Air Temperature is 15°C above set pointpointpoint | ||||
| Low Air Temperature: Activated if displayed Air Temperature is 25°C below set pointpointpoint | |||||
| Over Temperature Alarm | Over Air Temperature: Activated when displayed Air Temperature was approach 38°C in Air Mode Mode Mode | ||||
| System Alarm | System Failure: Activated if Temperature Control System malfunctions | ||||
| Temperature Control Range and Correlative Specifications | Air Temperature | 20°C~37°C~37°Ceg;C~37°C | |||
| Temperature Rise Time | ≤30min0min0min | ||||
| Temperature Variability | ±0.5°Cmn;0.5°Ceg;Cmn;0.5°C | ||||
| Temperature Uniformity | ≤1°C;1°Ceg;C;1°C | ||||
| Environment Requirement | Environment Temperature | Operation Temperature Range | +20°C~+30°C+30°Ceg;C+30°C | ||
| Store Temperature Range | -40°C~+55°C+55°Ceg;C+55°C | ||||
| Environment Humidity | Operation Humidity Range | 30%~75% | |||
| Store Humidity Range | 10%~100% | ||||
| Atmospheric Pressure Range | 700hPa~1060hPa | ||||
| Other Specifications | Noise Level Within Hood Environment | ≤55dB(Environment noise below 45dB)5dB)5dB) | |||
| Carbon Dioxide (CO2) Level within the Hood | Less than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress | ||||
| Air Velocity Over Mattress | less than 0.3m/s | ||||
| Infant Bed Max Load | 10kg | ||||
| Transfusion Shelf Max Load | 2kg | ||||
| Factory Default Settings | Temperature Control Mode Air Temperature Mode | ||||
| Air Set Temperature: 32°Cdeg;Cdeg;C | |||||
| Temperature Units:°Cdeg;Cdeg;C | |||||