Clinical Infant Radiant Warmer Medical Equipment (HNT-2000A)

Clinical Infant Radiant Warmer Medical Equipment (HNT-2000A)

1500-1600 USD

add to cart
Model NO. HNT-2000A
Certification CE, ISO13485
Medical Device Regulatory Type Type 2
Temperature Control Mode Pre-Heating Control, Servo Control & Manual ControControontro
Apgar Timer Built-in Apgar Timer and ﹥37°c Preset Function Functionction
Skin Sensor Precision ≤0.3°c0.3°cg;c0.3°c
Bed Tiltable Continues Adjustable in ±10°lusmn;10°eg;lusmn;10°
Ambit Environment Stream Velocity ≤0.3m/S.3m/S3m/S
Trademark Healicom
Origin China
Theory Infant Radiant Warmer
Group Infant
Medical Devices Reg./Record No. Jyjxz20122260356
Instant Parameter Display Set Temp, Skin Temp, Heat Power Rate;
Skin Temperature Sensor Measure Range 20°c~42°cc~42°cg;cc~42°c
Bed Surface Temperature Uniformity ≤1°ce;1°cg;ce;1°c
Operate Temperature Range 18°c~30°cc~30°cg;cc~30°c
Add Intra-Low-Pressure Suction Machine
Specification ISO
HS Code 90189090
Standard composition:
1. Temperature Control Mode: pre-heating control, servo control & manual control;trol;trol;
2. Instant Parameter Display: set temp, skin temp, heat power rate;
3. Built-in APGAR Timer and ﹥37°C preset functionfunctionction
4. Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
5. Tilting Bed with Rotary Radiant Head;
6. X-ray Cassette Tray under infant bed, examination light of high illuminance;
7. Use anti-blast quartz infrared tube as radiant heating source.
Technical specification:
Power AC 110 / 220 V, 50Hz / 60Hz
Input Power ≤1000VA00VA00VA
Temperature control range and relative specification Control mode Pre-warm mode, manual mode, baby mode
Temperature control range 25°C~37°C, for set temperature ≤37°Cure ≤37°C7°Cure ≤37°C
37°C~38°C, for set temperature ﹥37°Cure >37°Ceg;Cure >37°C
Skin temperature sensor measure range 20°C~42°C~42°Ceg;C~42°C
Deviation between sensor measure temperature and control temperature ≤0.5°C.5°Ceg;C.5°C
Skin Sensor Precision ≤0.3°C.3°Ceg;C.3°C
Bed surface temperature Uniformity ≤1°C;1°Ceg;C;1°C
Temperature Rise Time ≤45min5min5min
Bed tiltable continues adjustable in ±10°usmn;10°deg;usmn;10°
Alarm Sensor failure alarm Alarm with audible and visual when sensor open circuit, short circuit or disconnect, cut off power.
Over temperature alarm Alarm with audible and visual when displayed temperature was approach 38°C for set temperature ≤37°C, or approach 39°C for set temperature ﹥37°C, cut off power.7°C, cut off power..7°C, cut off power.
Deviation alarm After temperature equilibrium, alarm with audible and visual when displayed temperature was 1°C above or below control temperature, cut off power if above 1°C. 1°C.g;C. 1°C.
Power failure alarm Alarm with audible and visual when power failure
APGAR time Alarm with audible when run between 50 seconds to 1 minute, 4 minutes 50 seconds to 5 minutes, 9 minutes 50 seconds to 10 minutes.
Environment Temperature (It is a suggestion that please do not use outside of the allow environment.) Operate temperature range 18°C~30°C~30°Ceg;C~30°C
Store temperature range -20°C~+55°C+55°Ceg;C+55°C
Environment Humidity Operate humidity range 30%RH~75%RH
Store humidity range ≤93%RH3%RH3%RH
Atmospheric pressure Transport and Store atmospheric pressure range 500hPa~1060hPa
Operate atmospheric pressure range 700hPa~1060hPa
Stream velocity Ambit environment stream velocity ≤0.3m/s3m/s3m/s
Based on HNT-2000A, Add: Intra-low-pressure suction machine