Standard composition
1. Temperature Control Mode: pre-heating control, servo control & manual control;trol;trol;
2. Instant Parameter Display: set temp, skin temp, heat power rate;
3. Built-in APGAR Timer and ﹥37°C preset functionfunctionction
4. Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
5. Tilting Bed with Rotary Radiant Head;
6. X-ray Cassette Tray under infant bed, examination light of high illuminance;
7. Use anti-blast quartz infrared tube as radiant heating source.
Technical specification
| Power | AC 110 / 220 V, 50Hz / 60Hz |
| Input Power | ≤1000VA00VA00VA |
| Temperature control range and relative specification | Control mode | Pre-warm mode, manual mode, baby mode |
| Temperature control range | 25°C~37°C, for set temperature ≤37°Cure ≤37°C7°Cure ≤37°C |
| 37°C~38°C, for set temperature ﹥37°Cure >37°Ceg;Cure >37°C |
| Skin temperature sensor measure range | 20°C~42°C~42°Ceg;C~42°C |
| Deviation between sensor measure temperature and control temperature | ≤0.5°C.5°Ceg;C.5°C |
| Skin Sensor Precision | ≤0.3°C.3°Ceg;C.3°C |
| Bed surface temperature Uniformity | ≤1°C;1°Ceg;C;1°C |
| Temperature Rise Time | ≤45min5min5min |
| Bed tiltable | continues adjustable in ±10°usmn;10°deg;usmn;10° |
| Alarm | Sensor failure alarm | Alarm with audible and visual when sensor open circuit, short circuit or disconnect, cut off power. |
| Over temperature alarm | Alarm with audible and visual when displayed temperature was approach 38°C for set temperature ≤37°C, or approach 39°C for set temperature ﹥37°C, cut off power.7°C, cut off power..7°C, cut off power. |
| Deviation alarm | After temperature equilibrium, alarm with audible and visual when displayed temperature was 1°C above or below control temperature, cut off power if above 1°C. 1°C.g;C. 1°C. |
| Power failure alarm | Alarm with audible and visual when power failure |
| APGAR time | Alarm with audible when run between 50 seconds to 1 minute, 4 minutes 50 seconds to 5 minutes, 9 minutes 50 seconds to 10 minutes. |
| Environment Temperature (It is a suggestion that please do not use outside of the allow environment.) | Operate temperature range | 18°C~30°C~30°Ceg;C~30°C |
| Store temperature range | -20°C~+55°C+55°Ceg;C+55°C |
| Environment Humidity | Operate humidity range | 30%RH~75%RH |
| Store humidity range | ≤93%RH3%RH3%RH |
| Atmospheric pressure | Transport and Store atmospheric pressure range | 500hPa~1060hPa |
| Operate atmospheric pressure range | 700hPa~1060hPa |
| Stream velocity | Ambit environment stream velocity | ≤0.3m/s3m/s3m/s |
Add: Intra-low-pressure suction machine, white light photothreaphy unit, baby resuscitation bag, 7Loxygen cylinder*2, oxygen pressure meter and flow meter.