My-D012 100mA Medical X-ray Equipment X Ray Machine

My-D012 100mA Medical X-ray Equipment X Ray Machine

5700-6700 USD

add to cart
Model NO. MY-D012
Type X Ray Equipment
Medical Devices Reg./Record No. Jyjxzz20062300402
Maximum Tube Current 100mA
Transport Package Export Wooden Box
Classification Imaging Diagnostic Equipment
Group All
Maximum Tube Voltage 90kv
Trademark MAYA
Origin China
100mA Medical X-ray equipment(KH100 ), X-ray equipment for hospitals, x ray unit

We build good relationship and work strictly with manufacturers on quality inspection before shipment. The wide range of services we provide can reduce the number of suppliers you use, increasing your efficiency and cutting costs to your bottom line. Maya works for ALL your sourcing needs. We believe 'Customers' success is our success, and our success is customers' success'. It will be our great honor to serve you, Hope we can build up mutual business cooperation in the near future.near future.
MY-D012 100mA Medical x-ray equipment
Unit composition: Control Panel → high-voltage transformer component → high-voltage cable → X-ray tube components → diagnosis bed / gastrointestinal photographic devicel photographic devicel photographic device
Item Technical parameters
Power Supply
Voltage a. c380v/220v
Frequency 50Hz± 1Hzsmn; 1Hzsmn; 1Hz
Capacity 10Kva
Resistance 380v: 2Ω , 220 v: 0.6Ωv: 0.6ΩOmega;v: 0.6Ω
Maximum tube voltage 90KV
Maximum tube current 100mA
Fluoroscopy
Tube voltage 50~90kv
Tube current 0.1~5mA
Radiograph
Tube voltage 50~90kv
Tube current Small focus: 25Ma; 50, 75, 100mA divided
Spot film 75mA
Time 0.04 ~8.0S totally 23 grades
High-voltage generator Capacity 10KVA
X-ray Tube Model
Focus Small focus 1.8× 1.8mm2; Big focus4.3× 4.3 mm2imes; 4.3 mm2imes; 4.3 mm2
Diagnostic Bed
Rotating scope 90° ~0° ~-10°eg; ~-10°deg;eg; ~-10°
Line filter device Distance from the table≤ 60mm, f o=100cm00cm00cm
Limited beam splitter Light positioning limited beam splitter
Spot film device Spot film specification(gastrointestinal photograph) 128× 179mm(5' ' × 7' ' ) Half film' ) Half film' ) Half film
205× 254mm(8' ' × 10' ' ) Half film' ) Half film' ) Half film
254× 305mm(10' ' × 12' ' ) Half film' ) Half film' ) Half film
279× 355mm(11' ' × 14' ' ) A quarter of filmsrter of filmsrter of films
pillar Longitudinal movement(along the railway) 1400mm
Lateral movement 740mm
Up and down 700mm~1770mm(focus to ground)
Lock mode Electromagnetic Lock