Animal Applicable Multi-Channel Patient Monitor (BMO210V)

Animal Applicable Multi-Channel Patient Monitor (BMO210V)

bargain

add to cart
Model NO. BMO210V
Type Patient Monitor
Medical Device Regulatory Type Type 2
Product Size 32cm × 16cm × 27cmm × 27cm 27cmm × 27cm
Gross Weight 9.00kg
Display 12.1′′ LCDme;′ LCD; LCDme;′ LCD
Trademark BMV
Origin Shenzhen China
Classification Physiological Functions of Diagnosis and Monitoring Equipment
Group Animal
Medical Devices Reg./Record No. Yssyjx20150068
Product Weight 4.50kg
Package Size 32cm*16cm *27cm
Measuring Range -2.0mv ~ + 2.0mv
Specification CE
HS Code 9018193000
BMO210V Patient Monitor can monitor physiological parameter s: oxygen, ECG, body temperature, pulse, heart rate, blood pressure and so on. Reliable quality and stable health care system make policy makers more confidence . P owerful storage capacity, to ensure that you will not miss any clinical data.
BMO210V Patient Monitor applicable to all types of dogs and cats in clinical care, vital signs parameters more comprehensive and more scientific.
Features
    12.1'' color TFT LCD screen with multi-channel wave forms display
    Have patented technology of blood pressure measurement l
    Anti-defibrillation, anti electric knife, ECG digital filtering
    Super ECG anti-jamming capability
    English and other 14 languages Operating System
    Above 8000 Blood pressure storage and playback
    Above 680 hour trend graph subsisting and playback
    200 set alarm event storage and playback
    12 hour wave line storage and playback
    Built-in high-capacity rechargeable battery, with both AC and DC
    Can be connected to a central monitoring system,supports wired and wireless , system software support network upgrade
    Having a drug concentration calculations, breathing oxygen picture, ST segment analysis, large character interface, ECG full lead showed, abnormal heart rate analysis
    Detection parameters: ECG / heart rate, noninvasive blood pressure / pulse rate, oxygen saturation, respiration, pulse, temperature

1.ECG Specifications
Lead : three-lead or five-lead
Three leads: RA, LA, LL, Mode :I, II, III
Five leads: RA, LA, LL, RL, V, Mode : I, II, III, aVR, aVL, aVF, V
Gain : ´´acute;´ 250, ´´acute;´ 500, ´´acute;´ 1000, ´´acute;´ 2000
Heart Rate
Heart rate range :
Large dogs: 15 ~ 300bpm (stroke / min)
S mall dogs and puppies: 15 ~ 350 bpm (stroke / min)
Accuracy: ± 1% or ± 1bpm, whichever is greaterever is greaterever is greater
Resolution: 1 bpm (stroke / min)
Sensitivity: ﹥ 200 uV (peak to peak)eak)
Input impedance: ﹥ 5 (megohms)hms)
B andwidth
Diagnostic Mode: 0.05 ~ 130Hz
Monitoring mode: 0.5 ~ 40Hz
Operation mode: 1 ~ 20Hz
CMRR
Diagnostic mode:﹥ 90 dB0 dB
Monitoring mode:﹥ 100 dB0 dB
Operation mode:﹥ 100 dB0 dB
Electrode polarization voltage range : 300mV
Pacing pulse detection
To meet the following conditions pacing pulse can be detected:
Margin: ± 2 mV ~ ± 700mV ± 700mV; 700mV ± 700mV
Width: 0.1ms ~ 2ms
Rise time: 10us ~ 100μsmu;smu;s
Pacing pulse suppression
Margin: ± 2 mV ~ ± 700mV ± 700mV; 700mV ± 700mV
Width: 0.1ms ~ 2ms
Rise time: 10us ~ 100μsmu;smu;s
Baseline recovery time : after defibrillation <3 secondsonds
Signal Range : 8 mV (peak)
Calibration signal: 1mV (peak), accuracy 5%
ST segment detection measurements
Measuring range: -2.0mV ~ + 2.0mV
Accuracy: -0.8 mV ~ + 0.8mV ( error within ± 0.02mV or ± 10%or ± 10%mn; 10%or ± 10% )
Other ranges are not defined.
2. Respiratory Specifications
M easurement method : RA-LL impedance
Respiratory impedance detection range : 0.3 ~ 3Ω3Ωmega;3Ω
Base impedance range : 200 ~ 4000ΩΩOmega;Ω
Bandwidth : 0.1 ~ 2.5Hz
Respiration rate
Range: Large dogs: 2 ~ 120BrPM small dogs and puppies: 2 ~ 150 BrPM
Resolution: 1 BrPM
Accuracy: 2 BrPM
Suffocation alarm : 10 to 40 seconds
3. NIBP Specifications
M easurement method: Pulse wave oscillation
Work Mode : manual / automatic / STAT
M easurement interval for Automatic measurement mode 1,2,3,4,5,10,15,30,60,90,120,180,240,480 minutes
STAT measurement mode : measurement time 5 minutes
4. SpO2 Specifications
Oximetry
Measuring range 0 to 100%
Resolution 1%
Accuracy of 70 to 100%: 2; 0% -69%: Not defined
Pulse rate
Measuring range 20 ~ 300bpm
Resolution 1bpm
Accuracy 3bpm
5. TEMP Specifications
Temperature sensor : YSI series, CYF series
Number of channels 2 channels
M easuring
Range 0 ~ 50C
Resolution 0.1C
Accuracy 0.1C (sensor error not included)
6. IBP Specifications (optional)
Number of channels 2 channels.
P ressure t itle ART, PA, CVP, RAP, LAP, ICP, P1, P2.
Pressure measurement range
ART 0 ~ 300 mmHg
PA -6 ~ 120 mmHg
CVP -10 ~ 40 mmHg
RAP -10 ~ 40 mmHg
LAP -10 ~ 40 mmHg
ICP -10 ~ 40 mmHg
P1, P2 -10 ~ 300 mmHg
Accuracy ± 1mmHg or ± 2%, the value is greater (not including sensor error)ng sensor error)g sensor error)
Pressure Sensor
Sensitivity 5 V / V / mmHg
Impedance range 300 ~ 3000ΩΩOmega;Ω