Features:
1. It can measure the concentrations of K+, Na+, Cl- and Ca++ and pH values.
2. Applies ISE (Ion Selection Electrode), which gives the instrument the advantages of minimum maintenance, easy assembly and disassembly, as well as stable and reliable measurements.
3. Working Modes:
- Normal mode: Automatic calibration; This mode
Adapts to large and middle scale hospital with large sample volume; Sleep functions which can save reagent.
- Saving mode: 24 hours reagents consumption is only 7ml when it does not measure. This mode. Adapts to hospital with small sample volume.
4. Each test channel can be closed separately, So it is convenient to do emergencymeasurement.
5.110mm× 60mm LCD display, automatic screensaver, friendly interface, easy to operate.perate.perate.
6. Mini thermal printer can print repeatedly at any moment. It is convenient to feed paper. And the print is very clear.
7. Revising datas automatically to ensure the accuracy of data.
8. It can save1500 test results, data will not be lost even if the power is cut off. Users can inquire or print at any moment.
9. The instrument is equipped with a RS232 serial port, which can be connected to computer. It may collect, manage and print measure result by relevant software.
10. It has electrode-checking function. Users can do corresponding maintenance according to checking result
11. Exempt of routine maintenance, this greatly lighten the burden of operators; In-side software self-checking can suggest users eliminate the malfunction in time.
12. Advanced and reliable fluid position examination technology can save reagent
13. Adopt built-in operating system leads to a reliable performance.
14. Users can modify each setting to meet different requirements
15. The quality-control inspecting function can track the accuracy of the instrument.
Sample: Serum, Whole Blood, Plasma, Diluted Urine
Sample Volume: 90UL
Testing Time: ≤ 50s 50s 50s
Temperature: 10º C-30º C, humidity: ≤ 85%dity: ≤ 85%85%dity: ≤ 85%
Power Supply Voltage: AC160V-245V 50± 1Hzsmn; 1Hzsmn; 1Hz
Volume: 260mm× 260mm× 340mm× 340mm 340mm× 340mm
Weight: 6Kg
Packing Details: 355mm× 350mm× 420mm N. W. 7kgs G. W. 11kgss G. W. 11kgss G. W. 11kgs
Reagents:
1. Calibration solution 500ml× 3imes; 3imes; 3
2. Slop solution 500ml
3. Saturated KCLsolution 10ml
4. Protein-removing solution 5ml
5. Electrode solution 10ml
| Test Object | Linearity | Repeatability(Cv) | Stability | Accuracy |
| K+ | ± 0.3mmol/L.3mmol/L.3mmol/L | ≤ 1.0 %.0 %.0 % | ± 0.2mmol/L/hmmol/L/hmmol/L/h | ± 0.3mmol/L.3mmol/L.3mmol/L |
| Na+ | ± 3.0mmol/L.0mmol/L.0mmol/L | ≤ 1.0 %.0 %.0 % | ± 2.0mmol/L/hmmol/L/hmmol/L/h | ± 3.0mmol/L.0mmol/L.0mmol/L |
| Cl- | ± 3.0mmol/L.0mmol/L.0mmol/L | ≤ 1.0 %.0 %.0 % | ± 2.0mmol/L/hmmol/L/hmmol/L/h | ± 3.0mmol/L.0mmol/L.0mmol/L |
| Ca++ | ± 0.2mmol/L.2mmol/L.2mmol/L | ≤ 2.0 %.0 %.0 % | ± 0.2mmol/L/hmmol/L/hmmol/L/h | ± 0.2mmol/L.2mmol/L.2mmol/L |
| pH | — —dash; —mdash;dash; — | ≤ 1.0 %.0 %.0 % | ± 0.02/h; 0.02/h; 0.02/h |
|
| Standard Ca++ and Total Ca++(calculating value) |