Veterinary Infant Incubator Model: Neocare A Brief introduction: 1)Air mode control, set temperature and air temperature are displayed separately; 2)Natural air flow humidity; 3)Failure alarm indication; 4)RS-232 connector 5)NeoCare A2, NeoCare A3 are configured with fixed Neonate Bilirubin Phototherapy Equipment model: BLT-90 with blue fluorescence lamps) 6)NeoCare A2, NeoCare A3 are configured with fixed Neonate Bilirubin Phototherapy Equipment model: BLT-90L (with LED blue lights) Specification: 1)Power supply: AC220V-230V, 50Hz 2)Input power: 850VA 3)Mode of control: Air mode control by micro-computer 4)Control range of air temperatur: 25º C-37º C-37º Cdm; C-37º C 5)Variability of temperature: ≤ 0.5º C5º Cdm; C5º C 6)Uniformity of temperature: ≤ 0.8º C8º Cdm; C8º C 7)Noise inside hood: ≤ 60dB(A)B(A)B(A) 8)Failure alarm: Over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on. Requirements of environment: Environment temperature: 20º C~30º C~30º Cdm; C~30º C (Set temperature should be higher 3º C than the environment temperature)ature)ature) Wind speed of environment: <0.3m/s3m/s Package: Each unit is packed as one carton: Size: 108.5mm x 64mm x 108mm (NeoCare A/NeoCare A2) 108.5mm x 64mm x 112.5mm (NeoCare A1/NeoCare A3) Gross Weight: 72Kg (for NeoCare A/NeoCare A1) 83Kg (for NeoCare A2/NeoCare A3) Standard configuration: NeoCare A: Main body, stand, temperature controller, baby compartment, I. V. Pole, shelf and so on. NeoCare A1: Based on NeoCare A, added with the side door NeoCare A2: Based on NeoCare A, added with the Neonate bilirubin phototherapy equipment (fixed model) NeoCare A3: Based on NeoCare A1, added with the Neonate bilirubin phototherapy equipment (fixed model)