Hospital Neonate Bilirubin Phototherapy Equipment: Standard configure:
Temperature controller, Up Cabinet, Cabinet, infant bed, I.V.Pole.
Technical specification:
Power supply : AC220V 50HZ Power input: 800VA
Temperature control range: 25~37
The margin between the indication temperature and the control temperature: ≤0.7;0.7;0.7
Uniformity of the surface temperature on bed: ≤3.5;3.5;3.5
Radiation of Blue light for upper radiant box:≥850μW/cm2u;W/cm2u;W/cm2
Radiation of Blue light for lower radiant box:≥700μW/cm2u;W/cm2u;W/cm2
Timer range of blue light:0~9999 hours ( Reset )
Over temperature alarm: ≤39.5, give the visible and audible alarmlarmlarm
Sensor Alarm: The leads of sensor plug is open circuit or close circuit.
Package: | Gross Weight | Up Cabinet: 70Kg , Cabinet: 50 Kg |
| Net Weight | Up Cabinet: 50Kg , Cabinet: 40 Kg |
| Measurement | Up Cabinet: 1000mm×700mm×805mmm×805mm;805mmm×805mm |
| Cabinet | 900mm×600mm×885mmm×885mm;885mmm×885mm |