MY-E010 Medical Multifunctional Anesthesia Machine Main body High-strength engineering plastic rack, light, Beautiful and corrosion resistant Scope of application adult, child Gas source O2: 0.27~0.55MPa N2O: 0.27~0.55MPa Flow meter O2: 0.1~1.0L/min 1.1~10L/min N20: 0.1~1.0L/min 1.1~10L/min O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration﹥25%;25%; When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.off. Flow rate of rapid oxygen supply 25~75L/min Low oxygen pressure alarm There will be sound alarm when the oxygen Pressure < 0.2MPa2MPa Vaporizer it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. Respiratory circuit working mode: All-close, semi-close, semi-open APL: 0.5~7 kPa Respiratory Bellows bellows for adults, bellows for children Tidal volume range: 0~1, 500 ml Ventilator Display mode High-definition 5.7' LCD screen displayisplay Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH Ventilation parameters Tidal volume 50~1500 ml Rate 2~99 bpm SIMV rate 2~20 bpm I: E 2: 1~1: 8 Inspiratory trigger pressure -1.0~2.0 kPa PEEP 0~2.0 kPa Pressure range 1.0~6.0 kPa SIGH 1.5 times the inspiratory time Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure Monitoring of oxygen concentration (21%~100%) Security Alarm System Oxygen Concentration alarm upper limit setting range (50%~100%) Low limit setting range (15%~50%) Airway pressure alarm upper limit setting range 1.0~6.0 kPa Low limit setting range 0.4~2.0 kPa Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min Low limit setting range 1.0~10L/min Sustained high-pressure alarm it will give alarm when stress have consistently Been higher than 2.5 kPa Suffocation alarm it will give sound and light alarm if there is No tidal volume input for 15 seconds Power alarm Power AC 220V± 10% 50Hz± 1Hz0Hz± 1Hzmn; 1Hz0Hz± 1Hz (UPS, Storage battery) (Mechanical Arms) (End-expiratory CO2 monitoring) *Items within brackets are optional. Our Advantage : 1: For all your inquires about us or our products , we will reply you in detail within 24 hours. 2: We own professional team have professional attitude to recommend products and services. 3:One year free warranty , after it is still providing after-sales service . 4: We offer OEM services , Can print your own logo on the products . 5:We have very experienced engineers , can help you better use our products .