| Special function | With drug library |
| Rate: | 50/60ml Syringe: 0.1ml/h~999.9ml/h (0.1ml/h step) 1,000ml/h~1,500ml/h (1ml/h step) 30ml Syringe: 0.1ml/h~900.0ml/h (0.1ml/h step) 20ml Syringe: 0.1ml/h~600.0ml/h (0.1ml/h step) 10ml Syringe: 0.1ml/h~300.0ml/h (0.1ml/h step) 5ml Syringe: 0.1ml/h~100.0ml/h (0.1ml/h step) |
| Flow Rate Accuracy: | Within ±3% (after correct calibration) |
| Mechanical precision: | Within ±2% |
| Bolus Rate: | 50/60ml Syringe: 1,200ml/h 30 ml Syringe: 720ml/h 20 ml Syringe: 480ml/h 10 ml Syringe: 240ml/h 5 ml Syringe: 100ml/h |
| Purge Rate: | 50/60ml Syringe: 1,500ml/h 30ml Syringe: 900ml/h 20ml Syringe: 600ml/h 10ml Syringe: 300ml/h 5 ml Syringe: 100ml/h |
| Volume Limit: | 0.1ml~9999.9ml |
| Total Injection Volume: | 0.1ml~9999.9ml (0.1ml step) |
| Occlusion: | High: 800mmHg ±200mmHg (106.7kPa±26.7kPa) Medium: 50/600mmHg ±100mmHg (66.7kPa±13.3kPa) Low: 300mmHg ±100mmHg (40.7kPa±13.3kPa ) |
| Alarms: | Injection soon finish, end of injection, occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose etc. |
| Power Source: | AC 100V~240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup |
| KVO rate: | 1ml/h |
| Fuse: | F1AL/250/60V, 2pcs inside |
| Power Consumption: | 30VA |
| IP Classification: | IPX4 |
| Equipment Classification: | Class II, internal power supply, Type CF |
| Operating Condition: | Ambient temperature: +5ºC~+40ºC; |
| Relative humidity: | 20~90% |
| Transport & Storage Condition: | Ambient temperature: -30ºC~+55ºC |
| Relative humidity: | ≤95% |
| Dimension: | 280mm (L) × 210mm (W) × 130mm (H) |
| Weight: | 2.2kg (net weight) |