Classification Class II, Type CF
Product Feature
1. Compatible with 10ml/20ml/50ml standard syringes as open system
2. Micro flow rate until 0.1ml/h with increment of 0.1ml to satisfy the requirement for Neonatal/Pediatric
3. Anti-bolus system to reduce significantly bolus after occlusion sudden release
4. Reminder alarm in case of forgetting to start the operation
5. Alarm volume adjustable on 3 levels
6. Wireless data transmission with infusion supervision system
Product Technical Parameter
| Operating Mode | Flow Rate Mode |
| Syringe Type | 10, 20, 50 ml |
| Preset Volume | 0.0-999.9ml |
| Accumulated Injection Volume | 0.1 - 999.9 ml |
| Flow Rate Range | 10 ml syringe: 0.1 ml/h - 100 ml/h (increment 0.1 ml/h) |
| 20 ml syringe: 0.1 ml/h - 200 ml/h (increment 0.1 ml/h) |
| 50 ml syringe: 0.1 ml/h - 500 ml/h (increment 0.1 ml/h) |
| Accuracy | ± 3% |
| Bolus Rate | 10 ml syringe: 100 ml/h |
| 20 ml syringe: 200 ml/h |
| 50 ml syringe : 500 ml/h |
| KVO Rate | 0.1-5.0ml/h adjustable, 1ml/h as default |
Display
| Screen | 2.0 inch B/W LCD |
| Information | Flow Rate, Accumulated injection volume, Preset volume, Syringe size, Battery capacity, Bed No., AC power indicator, Battery charge indicator, etc. |
Alarm Type
1. Finish, Near, Syringe Empty, Syringe Disengaged, Occlusion, System Error, Battery Empty, Low Battery, Reminder to operate, No AC Power.
2. Occlusion Alarm Pressure Range 3 levels: 70kPa, 100kPa, 130kPa (±30kPa), Level Low as default.
Environment
| |
| Temperature 5 - 40 ºC for operating; -20 - 50 ºC for storage |
| Air Pressure 86 - 106 kPa for operating; 50 - 106 kPa for storage |
| Humidity 20 % - 80 % for operating; 10 % - 95 % for storage |