Ce Approved Medical Syringe Pump X-Pump S8

Ce Approved Medical Syringe Pump X-Pump S8

2+
1-500 USD

add to cart
Model NO. X-Pump S8
Type Syringe Pump
Group All Patient
Display Screen 2.0 Inch B/W LCD
Operating Mode Flow Rate Mode
Preset Volume 0.0-999.9ml
Kvo Rate 0.1-5.0ml/H Adjustable, 1ml/H as Default
Power Supply 100 - 240 VAC, 0.25-0.11A, 50/60 Hz
Trademark BMV
Specification 310 mm(L) × 125 mm(D) × 115 mm(H)mes; 115 mm(H)es; 115 mm(H)
HS Code 9018310000
Classification Syringe Pump
Certification CE, ISO13485
Medical Device Regulatory Type Type 2
Battery Lithium Polymer Battery
Syringe Type 10, 20, 50 Ml
Accumulated Injection Volume 0.1 - 999.9 Ml
Weight Approximately 1.8 Kg
Waterproof Level IP23
Transport Package Carton Box
Origin Shenzhen, China
X-Pump S8 Syringe Pump
designed for neonatal, pediatric and adult patients in the ICU and OR.
Product Features
Drug library with all the most used drugs
History record review and last configuration memory
Change the flow rate while operating
Manual bolus and automatic bolus
Anti-bolus system to reduce significantly bolus after occlusion sudden release
Micro flow rate of 0.1ml/h and Body weight mode to satisfy the requirement for Neonatal
Wireless data transmission with infusion supervision system (optional)
Specification list
X-Pump S8 Technical Specifications
Operating Mode Flow Rate Mode
Syringe Type 10, 20, 50 ml
Preset Volume 0.0-999.9ml
Accumulated Injection Volume 0.1 - 999.9 ml
Flow Rate Range 10 ml syringe: 0.1 ml/h - 100 ml/h (increment 0.1 ml/h)
20 ml syringe: 0.1 ml/h - 200 ml/h (increment 0.1 ml/h)
50 ml syringe: 0.1 ml/h - 500 ml/h (increment 0.1 ml/h)
Accuracy ± 3%usmn; 3%usmn; 3%
Bolus Rate 10 ml syringe: 100 ml/h
20 ml syringe : 200 ml/h
50 ml syringe : 500 ml/h
KVO Rate 0.1-5.0ml/h adjustable, 1ml/h as default
Display
Screen 2.0 inch B/W LCD
Information Flow Rate, Accumulated injection volume, Preset volume, Syringe
size, Battery capacity, Bed No., AC power indicator, Battery charge indicator,etc.
Alarm Type
Finish, Near, Syringe Empty, Syringe Disengaged, Occlusion, System Error, Battery Empty, Low Battery, Reminder to operate, No AC Power
Occlusion Alarm Pressure Range 3 levels: 70kPa, 100kPa, 130kPa (±30kPa), Level Low as default default default
Battery
Type Lithium polymer battery
Operation Time Up to 4 hours continuous operation @5 ml/h flow rate with standard battery
Up to 5 hours continuous operation @5 ml/h flow rate with optional battery
Environment
Temperature 5 - 40 ºC for operating, -20 - 60 ºC for storagefor storagefor storage
Air Pressure 57 - 106 kPa for operating, 50 - 106 kPa for storage
Humidity 15 % - 95 % for operating, 10 % - 95 % for storage
Others
Classification Class II, Type CF
Dimension 310 mm(L) × 125 mm(D) × 115 mm(H)es; 115 mm(H)es; 115 mm(H)
Weight Approximately 1.8 kg
Power Supply 100 - 240 VAC, 0.25-0.11A, 50/60 Hz
Waterproof level IP23