| Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant | ||
| Scope of application | adult, child | ||
| Gas source | O2: 0.27~0.55MPa N2O: 0.27~0.55MPa | ||
| Flow meter | O2: 0.1~1.0L/min 1.1~10L/min | ||
| N20: 0.1~1.0L/min 1.1~10L/min | |||
| O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration﹥25%; When the oxygen pressure﹤0.2Kpa, the flow of nitrous oxide would be cut off. | |||
| Flow rate of rapid oxygen supply | 25~75L/min | ||
| Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure ﹤ 0.2MPa | ||
| Vaporizer | it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. | ||
| Respiratory circuit | working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa | ||
| Respiratory Bellows | bellows for adults, bellows for children tidal volume range:0~1,500 ml | ||
| Ventilator | |||
| Display mode | High-definition TFT color LCD screen display | ||
| Ventilation mode | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, (IMV, VCV, PCV) | ||
| Ventilation parameters | Tidal volume | 50~1500 ml (20~1500 ml) | |
| Rate | 1~99 bpm | ||
| SIMV rate | 2~20 bpm | ||
| I:E | 2:1~1:8 (4:1~1:8) | ||
| Inspiratory trigger pressure | -1.0~2.0Kpa | ||
| PEEP Pressure Range | 0~2.0Kpa 1.0~6.0Kpa | ||
| SIGH | 1.5 times the inspiratory time (70-150 adjustable) | ||
| Inspiratory platform | 0~50% | ||
| Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow-volume oop,pressure-volume loop) | |||
| Monitoring of oxygen concentration | 21%~100% | ||
| Security Alarm System | |||
| Oxygen Concentration alarm | upper limit setting range 21%~100% | ||
| low limit setting range 10%~80% | |||
| Airway pressure alarm | upper limit setting range 0.3~6.0 kPa | ||
| low limit setting range 0.2~5.0 kPa | |||
| Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min | ||
| low limit setting range 1.0~10L/min | |||
| Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa | ||
| Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds | ||
| Power alarm | |||
| Power | AC 220V±10% 50Hz±1Hz(UPS, Storage battery) | ||
| (Mechanical Arms) (End-expiratory CO2 monitoring) | |||
| Items within brackets are optional. | |||