Air oxygen blender
KL-20 Features:
1. Application range: Recovery capsule oxygen, hood and masks oxygen, nasal pipeline,
drivepipe oxygen, warm box oxygen, NCPAP and ECMO.
2. Usage: In order to prevent POP for the premature and neonates, hood and nasal pipe and warm box was provided for oxygen..Air mixer type: Full mechanical air-oxygen mixer
3. Air operated and control, oxygen concentration and flow rate set separately and oxygen concentration and its direct parameters are adjustable.
4. Use special alloy materials and surface was anodizing, there is no oxygen battery and no internal
Power
5. Air source power: 0.28MPa ~ 0.6MPa
6. Imported GENTEC brand air input port
7. Oxygen concentration range: 21%~100%, accuracy: <±3%;±3%
8. Flow meter: GENTEC brand, 0~15LPM as basic function, 0-30LPM as optional function
9. Filter: sintered copper
10. Filter aperture: 75um Alarm: air source pressure difference 1.41bar±10% alarm(mechanical alarm). When the air difference reach to 1.41bar±10% alarm will ring.larm will ring.larm will ring.