Product parameters:
- Brand: Other
- Model: YD
- Customization: Yes
- Type: Digital Clamp Table
- Measuring voltage range: 20-500
- Measuring current range: 200ma-10
- Measuring resistance range: 40k
- Battery: A single 9V laminated battery with a supply current of less than 5mA
- Clamp opening size: Φ7mm × 9mm
First, the product overview
PY2000B double-clamp digital phase voltammeter is a high-precision, low-cost, hand-held, dual-channel input measuring instrument designed for field measurement of voltage, current and phase. The table can be easily measured in the field UU , II and UI between the phase, to determine the inductive, capacitive circuit and three-phase voltage phase sequence, the detection of the transformer wiring group, test the secondary circuit and bus differential protection system, read the differential protection between the groups CT Of the phase relationship, check the watt meter wiring is correct or not, etc. The use of clamp current transformer conversion method input current measured, so the measurement without disconnecting the measured line. U1-U2 measured between the phase, the two input circuit Completely insulated and isolated, so completely avoid the possibility of mistaken wiring caused by the measured circuit short circuit, so burned measuring instruments. Display using a high contrast LCD screen, the word up to 25mm, the screen angle can be freely converted about 70 ° to get the most Good visual effect.
Double-clamp digital phase voltammeter shell with engineering insulation material, reprovision rubber protective cover, safe and reliable.
Product name
Phase table, double clamp phase table, double clamp phase voltammeter, clamp phase table, clamp phase voltammeter, digital double clamp phase table, digital double clamp phase table, low voltage voltammetry phase detection table, handheld digital Double clamp phase voltammeter, double clamp digital phase table, double clamp digital phase table, double clamp type voltammetry phase table, double clamp voltamp phase table, digital double clamp phase table
two, Basic error
2.1 Reference working conditions
(A) Ambient temperature: (23 ± 5) ° C
(B) Ambient humidity: (45 to 75)% RH
(C) Measured signal waveform: sine wave, β = 0.02
(D) Measured signal frequency: (50 ± 0.2) Hz
(E) the position of the measured current conductor in the jaws: any
(F) When measuring the phase, the measured signal amplitude range: 100 ~ 220V, 0.5A ~ 1.5
(G) External reference frequency Electromagnetic field interference: should be avoided
2.2 Basic error limits
2.2.1 AC voltage (see Table 1)
Table 1: AC voltage measurement error |
Limit | Resolution | Basic error limit |
20V | 0.01V | ± (1.2% RD + 2) |
200V | 0.1V | ± (1.0% RD + 2) |
500V | 1V | ± (1.2% RD + 2) |
Input impedance: the limit is 2MΩ
2.2.2 AC current (see Table 2)
Table 2: AC current measurement error |
Limit | Resolution | Basic error limit |
200mA | 0.1mA | ± (1.0% RD + 2) |
2A | 1mA |
10A | 10mA |
2.2.3 phase
U-U, U-I, I-I (see Table 3)
Table 3: Power frequency phase measurement error |
Range | Resolution | Basic error limit |
0 ~ 360 ° | 1 | ± 3 ° |
When measuring U1-U2 phase voltage input circuit impedance: 40KΩ
three, Work error
3.1 Rated working conditions
(A) Ambient temperature: (0 to 40) ° C
(B) Ambient humidity: (20 to 80)% RH
(C) Measured signal waveform: sine wave, β = 0.05
(D) Measured signal frequency: (50 ± 0.5) Hz
(E) the position of the measured current conductor in the jaws: any
(F) the amplitude of the measured signal when the phase is measured
When measuring U1-U2 phase: 30V ~ 500V
When measuring I1-I2 phase: 10mA ~ 10.00A
When measuring U1-I2 or I1-U2 phase: 10V ~ 500V, 10mA ~ 10.00A
(G) External reference frequency Electromagnetic field interference: should be avoided
3.2 Rated working error limit
Under the rated operating conditions described in 2.1, the rated operating error limits shall not exceed twice the corresponding basic error limits.
four, Other technical features
4.1 Display the number of digits: three and a half
4.2 Sampling rate: 3 times / sec
4.3 Power supply: a single 9V laminated battery, the supply current is less than 5mA
4.4 Dimensions
Case size: 192mm × 95mm × 55mm
Shell size: 140mm × 42mm × 20mm
Jaw size: Φ7mm × 9mm
4.5 weight
Body: 280g
Measuring pliers: 2 x 200 g
4.6 Storage conditions
Temperature: - 10 ℃ ~ 50 ℃