Manufacturers of intelligent insulation resistance tester | digital insulation resistance tester | megger

Manufacturers of intelligent insulation resistance tester | digital insulation resistance tester | megger

1+
5000 CNY (728.02USD)
52+
4800 CNY (698.9USD)

add to cart

Product parameters:

  • Customization: Yes
  • Type: digital resistance measuring instrument
  • Brand: Yu De
  • Model: YD
  • Measuring range: 0-20000
  • Power supply: 220
  • Test voltage: 5000V-10K
  • Accuracy: 5% ± 10V
  • Repeat error: ± (10% + 5 words)
  • Ambient temperature: -10 ~ 40 ℃
  • Relative humidity: 20 to 80%
  • Calibration cycle: 1
  • Weight: 2
  • Size: 3




Insulation resistance tester dedicated to the laboratory or on-site insulation test test. Contains high-precision micro-current measurement system, digital boost system. Only need to use a high voltage line and a signal line can be connected to the test sample measurement automatically, the results By the large-screen LCD display, and the results of storage.

First, the main features

1. Using 32-bit micro-controller control, the entire Chinese user interface, easy to operate.

2. Automatically calculate the absorption ratio and polarization index, and automatically store 15 seconds, 1 minute, 2 minutes, 10 minutes per minute data for easy analysis.

3. Output current is large, short circuit current is greater than 5mA.

4. Anti-interference ability, to meet the ultra-high voltage substation site operation.

5. Test the automatic discharge, and real-time monitoring of the discharge process.

6. Rechargeable battery and charger, when not using the backlight, fully charged can be used continuously for 6 to 12 hours.

Second, the main technical performance

Accuracy: ± (10% + 5 words)

Measuring range: 0.1M ~ 200GΩ

Test voltage: Setting range: 0.5Kv ~ 10KV Accuracy: 5% ± 10V

Step-up and down mode: continuous smooth adjustment

Short circuit current:> 5mA

Measurement time: 1 minute ~ 10 minutes (with the measurement method)

Charging power: 180 ~ 270VAC, 50Hz / 60Hz ± 1% (mains or generator power supply)

Working environment: temperature -10 ~ 40 ℃, relative humidity 20 ~ 80%.